Magnetic resonance imaging in patients with temporary external pacemakers
- PMID: 36006800
- DOI: 10.1093/europace/euac147
Magnetic resonance imaging in patients with temporary external pacemakers
Abstract
Aims: To describe safety and feasibility of magnetic resonance imaging (MRI) in patients with transvenous temporary external pacemakers and whether artefacts affect the diagnostic image quality during cardiac MRI.
Methods and results: We reviewed records of all patients treated with temporary external pacing between 2016 and 2020 at a tertiary centre. Temporary pacing was established using a transvenous standard active fixation pacing lead inserted percutaneously and connected to a MRI-conditional pacemaker taped to the skin. All patients undergoing cardiac or non-cardiac MRI during temporary transvenous pacing were identified. Before MRI, devices were programmed according to guidelines for permanent pacemakers, and patients were monitored with continuous electrocardiogram during MRI. Of 827 consecutive patients receiving a temporary external pacemaker, a total of 44 (5%) patients underwent MRI (mean age 71 years, 13 [30%] females). Cardiac MRI was performed in 22 (50%) patients, while MRI of cerebrum, spine, and other regions was performed in the remaining patients. Median time from implantation of the temporary device to MRI was 6 (3-11) days. During MRI, we observed no device-related malfunction or arrhythmia. Nor did we detect any change in lead sensing, impedance, or pacing threshold. We observed no artefacts from the lead or pacemaker compromising the diagnostic image quality of cardiac MRI. MRI provided information to guide the clinical management in all cases.
Conclusion: MRI is feasible and safe in patients with temporary external pacing established with a regular MRI-conditional pacemaker and a standard active fixation lead. No artefacts compromised the diagnostic image quality.
Keywords: Diagnostic imaging quality; Feasibility; Magnetic resonance imaging; Safety; Temporary external pacemakers.
© The Author(s) 2022. Published by Oxford University Press on behalf of the European Society of Cardiology. All rights reserved. For permissions, please email: journals.permissions@oup.com.
Conflict of interest statement
Conflict of interest: C.T.W. is supported by grants from Biotronik. All others declare no conflicts of interest.
Similar articles
-
Feasibility of cardiac magnetic resonance imaging in temporary permanent pacemaker implants in pediatric myocarditis and complete atrioventricular block.J Arrhythm. 2024 Feb 29;40(2):385-389. doi: 10.1002/joa3.13007. eCollection 2024 Apr. J Arrhythm. 2024. PMID: 38586858 Free PMC article.
-
An eight-year prospective controlled study about the safety and diagnostic value of cardiac and non-cardiac 1.5-T MRI in patients with a conventional pacemaker or a conventional implantable cardioverter defibrillator.Eur Radiol. 2018 Jun;28(6):2406-2416. doi: 10.1007/s00330-017-5098-z. Epub 2018 Jan 9. Eur Radiol. 2018. PMID: 29318430 Clinical Trial.
-
Safety and efficiency of low-field magnetic resonance imaging in patients with cardiac rhythm management devices.Eur J Radiol. 2019 Sep;118:96-100. doi: 10.1016/j.ejrad.2019.07.005. Epub 2019 Jul 9. Eur J Radiol. 2019. PMID: 31439265
-
A Review of Temporary Permanent Pacemakers and a Comparison with Conventional Temporary Pacemakers.J Innov Card Rhythm Manag. 2019 May 15;10(5):3652-3661. doi: 10.19102/icrm.2019.100506. eCollection 2019 May. J Innov Card Rhythm Manag. 2019. PMID: 32477730 Free PMC article. Review.
-
Insertion and Management of Temporary Pacemakers.Semin Cardiothorac Vasc Anesth. 2016 Mar;20(1):52-62. doi: 10.1177/1089253215584923. Epub 2015 Jun 30. Semin Cardiothorac Vasc Anesth. 2016. PMID: 26134176 Review.
Cited by
-
Cardiac computed tomography-verified right ventricular lead position and outcomes in cardiac resynchronization therapy.J Interv Card Electrophysiol. 2022 Sep;64(3):783-792. doi: 10.1007/s10840-022-01193-1. Epub 2022 Mar 28. J Interv Card Electrophysiol. 2022. PMID: 35348998 Review.
-
Feasibility of cardiac magnetic resonance imaging in temporary permanent pacemaker implants in pediatric myocarditis and complete atrioventricular block.J Arrhythm. 2024 Feb 29;40(2):385-389. doi: 10.1002/joa3.13007. eCollection 2024 Apr. J Arrhythm. 2024. PMID: 38586858 Free PMC article.
-
Interventional cardiovascular magnetic resonance: state-of-the-art.J Cardiovasc Magn Reson. 2023 Aug 14;25(1):48. doi: 10.1186/s12968-023-00956-7. J Cardiovasc Magn Reson. 2023. PMID: 37574552 Free PMC article. Review.
-
Temporary transvenous external pacing for cardiac MRI in a pediatric patient.Pediatr Radiol. 2025 Jul 31. doi: 10.1007/s00247-025-06325-z. Online ahead of print. Pediatr Radiol. 2025. PMID: 40742405
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical