Is pre-heat necessary for the measurement of 8-oxo-7,8-dihydroguanosine and 8-oxo-7,8-dihydro-2'-deoxyguanosine in urine samples
- PMID: 36036744
- PMCID: PMC9550956
- DOI: 10.1002/jcla.24674
Is pre-heat necessary for the measurement of 8-oxo-7,8-dihydroguanosine and 8-oxo-7,8-dihydro-2'-deoxyguanosine in urine samples
Abstract
Background: It is currently unclear for the necessary of pre-heating urine samples for the accurate determination of 8-oxo-7,8-dihydroguanosine (8-oxoG) and 8-oxo-7,8-dihydro-2'-deoxyguanosine (8-oxodG). Thus, we conducted this study to evaluate the effect of pre-heat (i.e., to 37°C) on the accurate measurement of 8-oxoG and 8-oxodG in frozen urine samples.
Methods: Random urine samples from six healthy volunteers, six patients with renal dysfunction, and six patients with systematic diseases such as diabetes were collected, split, and stored at -80°C for up to 1 month. The frozen samples were thawed at room temperature (RT) or 37°C for different time, 10-fold diluted with ddH2O containing 1% formic acid, and determined by self-established LC-MS/MS method coupled with an ACQUITY™ Primer HSS T3 column.
Results: Thawing the samples at RT for 30 or 120 min, or at 37°C for 15 or 90 min did not affect the determination of 8-oxoG and 8-oxodG in urine samples. Moreover, no significant difference between thawing the urine samples at RT and 37°C was found after storing at -80°C for 1-3 months.
Conclusion: It is not always necessary to pre-heat the frozen urine samples to release 8-oxoG and 8-oxodG from precipitates, which is associated with different pre-treatment and determination methods.
Keywords: 8-oxo-7,8-dihydro-2′-deoxyguanosine; 8-oxo-7,8-dihydroguanosine; LC-MS/MS; pre-heat.
© 2022 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC.
Conflict of interest statement
The authors declare that they have no competing interests.
Figures


Comment in
-
Commentary on: Is preheat necessary for the measurement of 8-oxo-7,8-dihydroguanosine and 8-oxo-7,8-dihydro-2'-deoxyguanosine in urine samples.J Clin Lab Anal. 2023 Mar;37(6):e24874. doi: 10.1002/jcla.24874. Epub 2023 Apr 1. J Clin Lab Anal. 2023. PMID: 37003618 Free PMC article. No abstract available.
Similar articles
-
Nightshift work can induce oxidative DNA damage: a pilot study.BMC Public Health. 2023 May 15;23(1):891. doi: 10.1186/s12889-023-15742-4. BMC Public Health. 2023. PMID: 37189122 Free PMC article.
-
A serially coupled stationary phase method for the determination of urinary 8-oxo-7,8-dihydro-2'-deoxyguanosine by liquid chromatography ion trap tandem mass spectrometry.Redox Biol. 2013 Oct 16;1(1):492-7. doi: 10.1016/j.redox.2013.10.001. eCollection 2013. Redox Biol. 2013. PMID: 24251117 Free PMC article.
-
A robust LC-MS/MS method to measure 8-oxoGuo, 8-oxodG, and NMN in human serum and urine.Anal Biochem. 2023 Jan 1;660:114970. doi: 10.1016/j.ab.2022.114970. Epub 2022 Oct 29. Anal Biochem. 2023. PMID: 36341768
-
8-oxoguanine and 8-oxodeoxyguanosine Biomarkers of Oxidative DNA Damage: A Review on HPLC-ECD Determination.Molecules. 2022 Mar 1;27(5):1620. doi: 10.3390/molecules27051620. Molecules. 2022. PMID: 35268721 Free PMC article. Review.
-
8-Oxo-7,8-Dihydro-2'-Deoxyguanosine (8-oxodG) and 8-Hydroxy-2'-Deoxyguanosine (8-OHdG) as a Potential Biomarker for Gestational Diabetes Mellitus (GDM) Development.Molecules. 2020 Jan 3;25(1):202. doi: 10.3390/molecules25010202. Molecules. 2020. PMID: 31947819 Free PMC article. Review.
Cited by
-
Nightshift work can induce oxidative DNA damage: a pilot study.BMC Public Health. 2023 May 15;23(1):891. doi: 10.1186/s12889-023-15742-4. BMC Public Health. 2023. PMID: 37189122 Free PMC article.
-
Commentary on: Is preheat necessary for the measurement of 8-oxo-7,8-dihydroguanosine and 8-oxo-7,8-dihydro-2'-deoxyguanosine in urine samples.J Clin Lab Anal. 2023 Mar;37(6):e24874. doi: 10.1002/jcla.24874. Epub 2023 Apr 1. J Clin Lab Anal. 2023. PMID: 37003618 Free PMC article. No abstract available.
-
Reply to commentary on "Is pre-heat necessary for the measurement of 8-oxo-7,8-dihydroguanosine and 8-oxo-7,8-dihydro-2'-deoxyguanosine in urine samples".J Clin Lab Anal. 2023 Mar;37(6):e24873. doi: 10.1002/jcla.24873. Epub 2023 Apr 9. J Clin Lab Anal. 2023. PMID: 37032429 Free PMC article. No abstract available.
References
-
- Hu CW, Wang CJ, Chang LW, Chao MR. Clinical‐scale high‐throughput analysis of urinary 8‐oxo‐7,8‐dihydro‐2′‐deoxyguanosine by isotope‐dilution liquid chromatography‐tandem mass spectrometry with on‐line solid‐phase extraction. Clin Chem. 2006;52(7):1381‐1388. - PubMed
-
- Weimann A, Belling D, Poulsen HE. Measurement of 8‐oxo‐2′‐deoxyguanosine and 8‐oxo‐2′‐deoxyadenosine in DNA and human urine by high performance liquid chromatography‐electrospray tandem mass spectrometry. Free Radic Biol Med. 2001;30(7):757‐764. - PubMed
-
- Henriksen T, Weimann A, Larsen EL, Poulsen HE. Quantification of 8‐oxo‐7,8‐dihydro‐2′‐deoxyguanosine and 8‐oxo‐7,8‐dihydro‐guanosine concentrations in urine and plasma for estimating 24‐h urinary output. Free Radic Biol Med. 2021;172:350‐357. - PubMed
-
- Shi F, Nie B, Gan W, et al. Oxidative damage of DNA, RNA and their metabolites in leukocytes, plasma and urine of Macaca mulatta: 8‐oxoguanosine in urine is a useful marker for aging. Free Radic Res. 2012;46(9):1093‐1098. - PubMed
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources