Operational Performance of a Compact Proton Therapy System: A 5-Year Experience
- PMID: 36060418
- PMCID: PMC9415750
- DOI: 10.14338/IJPT-21-00033.1
Operational Performance of a Compact Proton Therapy System: A 5-Year Experience
Abstract
Purpose: We present an analysis of various operational metrics for a novel compact proton therapy system, including clinical case mix, subsystems utilization, and quality assurance trends in beam delivery parameters over a period of 5 years.
Materials and methods: Patient-specific data from a total of 850 patients (25,567 fractions) have been collected and analyzed. The patient mix include a variety of simple, intermediate, and complex cases. Beam-specific delivery parameters for a total of 3585 beams were analyzed. In-room imaging system usage for off-line adaptive purpose is reported. We also report key machine performances metrics based on routine quality assurance in addition to uptime.
Results: Our analysis shows that system subcomponents including gantry and patient positioning system have maintained a tight mechanical tolerance over the 5-year period. Various beam parameters were all within acceptable tolerances with no clear trends. Utilization frequency histograms of gantry and patient positioning system show that only a small fraction of all available angles was used for patient deliveries with cardinal angels as the most usable. Similarly, beam-specific metrics, such as range, modulation, and air gaps, were clustered unevenly over the available range indicating that this compact system was more than capable to treat the complex variety of tumors of our patient mix.
Conclusion: Our data show that this compact system is versatile, robust, and capable of delivering complex treatments like a large full-gantry system. Utilization data show that a fraction of all subcomponents range of angular motion has been used. Compilation of beam-specific metrics, such as range and modulation, show uneven distributions with specific clustering over the entire usable range. Our findings could be used to further optimize the performance and cost-effectiveness of future compact proton systems.
Keywords: compact systems; operational performance; proton therapy.
©Copyright 2022 The Author(s).
Conflict of interest statement
Conflicts of interest: Omar A. Zeidan, PhD, serves on the technical advisory committee for Mevion Medical Systems. All other authors have no relevant conflicts of interest to disclose.
Figures
References
-
- Contreras J, Zhao T, Perkins S, Sun B, Goddu S, Mutic S, Bottani B, Endicott S, Michalski J, Robinson C, Tsien C, Huang J, Fischer-Valuck BW, Hallahan D, Klein E, Bradley J. The world's first single-room proton therapy facility: two-year experience. Prac Radiat Oncol . 2017;7:e71–e76. - PubMed
-
- Elnahal SM, Kerstiens J, Helsper RS, Zietman AL, Johnstone PA. Proton beam therapy and accountable care: the challenges ahead. Int J Radiat Oncol Biol Phys . 2013;85:e165–e172. - PubMed
-
- Meeks SL, Shah AP, Sood G, Dvorak T, Zeidan OA, Meeks DT, Kelly P. Effect of proposed episode-based payment models on advanced radiotherapy procedures. JCO Oncol Pract . 2020;16:e1943–e1948. - PubMed
-
- Suzuki K, Gillin MT, Sahoo N, Zhu XR, Lee AK, Lippy D. Quantitative analysis of beam delivery parameters and treatment process time for proton beam therapy. Med Phys . 2011;38:4329–4337. - PubMed
LinkOut - more resources
Full Text Sources