Optimization and validation of a fast supercritical fluid chromatography tandem mass spectrometry method for the quantitative determination of a large set of PFASs in food matrices and human milk
- PMID: 36115197
- DOI: 10.1016/j.jchromb.2022.123455
Optimization and validation of a fast supercritical fluid chromatography tandem mass spectrometry method for the quantitative determination of a large set of PFASs in food matrices and human milk
Abstract
An Ultra-High Performance Supercritical Fluid Chromatography coupled with tandem Mass Spectrometry analytical method (UHPSFC-MS/MS) was developed for the determination of 34 perfluoroalkylated substances (PFASs) in food-related matrices. Two parameters (i.e. stationary phase and co-solvent) were selected and optimized using a step-by-step method, while a design of experiment (DoE) method using a central composite design (CCD) was implemented to optimize column temperature, mobile phase flow rate, co-solvent concentration and automated back pressure regulator (ABPR). The Torus 2-PIC column was selected along with ammonium acetate AcoNH4 as additive in the co-solvent. DoE optimization of both peak width and resolution enabled validating an optimized model (desirability 0.613) and setting column temperature at 38.7 °C, AcoNH4 concentration at 8 mM, mobile phase flow rate of 1.9 mL/min and ABPR at 1654 psi. The validated resulting method enabled reaching limits of quantification below 0.2 ng/g (w.w.) for 97 % PFASs in accordance with current EU requirements. The strategy was successfully applied to the characterization of a range (n > 30) of food-related matrices (red meat, poultry meat, eggs, fish and breast milk) collected in Algeria in 2019. PFOA and PFBA were observed as the most frequently detected PFASs, i.e. in 96.96 % and 90.9 % of the samples respectively. The highest concentrations were determined in fishery products up to 4.42 ng/g (w.w.) for PFTeDA and 0.75 ng/g (w.w.) for PFOS.
Keywords: Food; Human milk; PFASs; Supercritical fluid chromatography; Tandem mass spectrometry.
Copyright © 2022 Elsevier B.V. All rights reserved.
Conflict of interest statement
Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources