Patient-Ventilator Synchrony
Abstract
Patient-ventilator asynchrony develops when the ventilator output does not match the efforts of the patient and contributes to excess work of breathing, lung injury, and mortality. Asynchronies are categorized as trigger (breath initiation), flow (delivery of the breath), and cycle (transition from inspiration to expiration). Clinicians should be skilled at ventilator waveform analysis to detect patient-ventilator asynchronies and make informed ventilator adjustments. Ventilator overdrive suppresses respiratory drive and reduces asynchrony, while other adjustments specific to the asynchrony are also useful.
Keywords: Asynchrony; Dysychrony; Mechanical ventilation; Respiratory failure.