Patient safety program and incident review of high-dose-rate brachytherapy at an academic center in Thailand
- PMID: 36199945
- PMCID: PMC9528833
- DOI: 10.5114/jcb.2022.118793
Patient safety program and incident review of high-dose-rate brachytherapy at an academic center in Thailand
Abstract
Purpose: An incident review of errors related to using high-dose-rate brachytherapy (HDR-BT) and associated patient safety program were presented. This study was based on 9 years' experience using VariSource afterloader system.
Material and methods: Analysis was made on radiotherapy (RT) incidents (including near-misses) that were routinely recorded using manual and electronic incident reporting systems between July 2012 and December 2021. Each incident's origin was categorized as 'apparatus', 'system functionality', 'treatment procedure', and 'other causes'.
Results: From 8,100 fractions and 2,216 patients, there were 164 RT incidents. The most frequent cases included non-dosimetric errors arising from system malfunction (49/151), difficulties caused by insufficient instruments (47/151), problems with treatment delivery (55/151), and planning procedure errors (13/142). Eleven incidents were near-misses, another 11 were not harmful, and zero were harmful. The frequency rate of dosimetric errors was 0.14 per 100 BT insertions, and 0.5 per 100 patients. The review also discovered 45 of the 164 incidents related to tube sensor failures and source blockages. These delivery errors were associated with 0.56 incidents per 100 insertions and 2.03 incidents per 100 patients, inconveniencing patients in treatment delays.
Conclusions: The effectiveness of our HDR-BT safety program was evidenced by low-rate of dosimetric errors. Based on the analysis of 9 years of incidents, the error sources included uncommon or complex procedures, human factors, and work environment (equipment availability and maintenance).
Keywords: RT events; high-dose-rate brachytherapy; incidents reporting; incidents review; patient safety.
Copyright © 2022 Termedia.
Conflict of interest statement
The authors report no conflict of interest.
Figures
References
-
- Nag S, Cardenes H, Chang Set al. . Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: Report from Image-Guided Brachytherapy Working Group. Int J Radiat Oncol Biol Phys 2004; 60: 1160-1172. - PubMed
-
- Haie-Meder C, Potter R, Van Limbergen Eet al. . Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): Concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol 2005; 74: 235-245. - PubMed
-
- Potter R, Haie-Meder C, Van Limbergen Eet al. . Recommendations from gynecological (GYN) GEC ESTRO Working Group (II): Concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy-3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol 2006; 78: 67-77. - PubMed
-
- Kim RY, Shen S, Duan J. Image-based three-dimensional treatment planning of intracavitary brachytherapy for cancer of the cervix: Dose-volume histograms of the bladder, rectum, sigmoid colon, and small bowel. Brachytherapy 2007; 6: 187-194. - PubMed
-
- Erickson BA, Demanes DJ, Ibbott GSet al. . American Society for Radiation Oncology (ASTRO) and American College of Radiology (ACR) practice guideline for the performance of high-dose-rate brachytherapy. Int J Radiat Oncol Biol Phys 2011; 79: 641-649. - PubMed
LinkOut - more resources
Full Text Sources