. 2022 Oct;38(10):648-653.
doi: 10.3928/1081597X-20220715-02. Epub 2022 Oct 1.

Effects of Modified Haptics on Surgical Outcomes and Rotational Stability of Toric Intraocular Lens Implantation

Effects of Modified Haptics on Surgical Outcomes and Rotational Stability of Toric Intraocular Lens Implantation

Ryoko Osawa et al. J Refract Surg. 2022 Oct.

Abstract

Purpose: To assess the rotational stability of a new toric intraocular lens (IOL), TECNIS toric II (toric II), which is a modified version of the TECNIS toric IOL (toric I) with frosted haptics (Johnson & Johnson).

Methods: A total of 101 eyes of 101 patients who had been treated with phacoemulsification and toric IOL implantation were included. Before and 1 day, 1 week, and 1 month after surgery, uncorrected (UDVA) and corrected (CDVA) distance visual acuity were measured. Preoperative corneal astigmatism and postoperative manifest refractive astigmatism at 1 day and 1 month were analyzed. At 1 day and 1 month postoperatively, the amount of IOL axis misalignment from the intended orientation, tilt, and decentration were measured using anterior segment optical coherence tomography.

Results: Fifty-one eyes received the toric I IOL and 50 eyes received the toric II IOL. Toric I IOLs showed a significantly larger amount of axis misalignment than toric II IOLs at both 1 day (9.6 ± 7.6° vs 5.4 ± 4.8°, P = .003) and 1 month (9.1 ± 7.8° vs. 4.7 ± 4.2°, P = .003) postoperatively.The proportion of eyes with misalignment greater than 10° was significantly larger with toric I than toric II IOLs (P < .001). There were no significant differences between IOLs in the amount of residual astigmatism, UDVA, CDVA, and amount of tilt and decentration at 1 day and 1 month postoperatively.

Conclusions: The TECNIS toric II IOL with frosted haptics has significantly improved rotational stability compared to its previous model. [J Refract Surg. 2022;38(10):648-653.].