Metal artifact correction in photon-counting detector computed tomography: metal trace replacement using high-energy data
- PMID: 36227611
- DOI: 10.1002/mp.16049
Metal artifact correction in photon-counting detector computed tomography: metal trace replacement using high-energy data
Abstract
Background: Metal artifacts have been an outstanding issue in computed tomography (CT) since its first uses in the clinic and continue to interfere. Metal artifact reduction (MAR) methods continue to be proposed and photon-counting detectors (PCDs) have recently been the subject of research toward this purpose. PCDs offer the ability to distinguish the energy of incident x-rays and sort them in a set number of energy bins. High-energy data captured using PCDs have been shown to reduce metal artifacts in reconstructions due to reduced beam hardening.
Purpose: High-energy reconstructions using PCD-CT have their drawbacks, such as reduced image contrast and increased noise. Here, we demonstrate a MAR algorithm, trace replacement MAR (TRMAR), in which the data corrupted by metal artifacts in full energy spectrum projections are corrected using the high-energy data captured during the same scan. The resulting reconstructions offer similar MAR to that seen in high-energy reconstructions, but with improved image quality.
Methods: Experimental data were collected using a bench-top PCD-CT system with a cadmium zinc telluride PCD. Simulations were performed to determine the optimal high-energy threshold and to test TRMAR in simulations using the XCAT phantom and a biological sample. For experiments a 100-mm diameter cylindrical phantom containing vials of water, two screws, various densities of Ca(ClO4 )2 , and a spatial resolution phantom was imaged with and without the screws. The screws were segmented in the initial reconstruction and forward projected to identify them in the sinogram space in order to perform TRMAR. The resulting reconstructions were compared to the control and to reconstructions corrected using normalized metal artifact reduction (NMAR). Additionally, a beef short rib was imaged with and without metal to provide a more realistic phantom.
Results: XCAT simulations showed a reduction in the streak artifact from -978 HU in uncorrected images to -10 HU with TRMAR. The magnitude of the metal artifact in uncorrected images of the 100-mm phantom was -442 HU, compared to the desired -81 HU with no metal. TRMAR reduced the magnitude of the artifact to -142 HU, with NMAR reducing the magnitude to -96 HU. Relative image noise was reduced from 176% in the high-energy image to 56% using TRMAR. Density quantification was better with NMAR, with the Ca(ClO4 )2 vial affected most by metal artifacts showing 0.8% error compared to 2.1% with TRMAR. Small features were preserved to a greater extent with TRMAR, with the limiting spatial frequency at 20% of the MTF fully maintained at 1.31 lp/mm, while with NMAR it was reduced to 1.22 lp/mm. Images of the beef short rib showed better delineation of the shape of the metal using TRMAR.
Conclusions: NMAR offers slightly better performance compared to TRMAR in streak reduction and image quality metrics. However, TRMAR is less susceptible to metal segmentation errors and can closely approximate the reduction in the streak metal artifact seen in NMAR at 1/3 the computation time. With the recent introduction of PCD-CT into the clinic, TRMAR offers notable potential for fast, effective MAR.
Keywords: computed tomography; metal artifact reduction; photon-counting detector CT.
© 2022 American Association of Physicists in Medicine.
Similar articles
-
Addressing CT metal artifacts using photon-counting detectors and one-step spectral CT image reconstruction.Med Phys. 2022 May;49(5):3021-3040. doi: 10.1002/mp.15621. Epub 2022 Apr 5. Med Phys. 2022. PMID: 35318699 Free PMC article.
-
A nonlinear scaling-based normalized metal artifact reduction to reduce low-frequency artifacts in energy-integrating and photon-counting CT.Med Phys. 2023 Aug;50(8):4721-4733. doi: 10.1002/mp.16461. Epub 2023 May 18. Med Phys. 2023. PMID: 37202918
-
Photon-counting normalized metal artifact reduction (NMAR) in diagnostic CT.Med Phys. 2021 Jul;48(7):3572-3582. doi: 10.1002/mp.14931. Epub 2021 May 28. Med Phys. 2021. PMID: 33973237
-
CT artifact correction for sparse and truncated projection data using generative adversarial networks.Med Phys. 2021 Feb;48(2):615-626. doi: 10.1002/mp.14504. Epub 2020 Dec 30. Med Phys. 2021. PMID: 32996149 Review.
-
Advances in metal artifact reduction in CT images: A review of traditional and novel metal artifact reduction techniques.Eur J Radiol. 2024 Jan;170:111276. doi: 10.1016/j.ejrad.2023.111276. Epub 2023 Dec 21. Eur J Radiol. 2024. PMID: 38142571 Review.
Cited by
-
Photon-counting Detector CT in Patients Pre- and Post-Transcatheter Aortic Valve Replacement.Radiol Cardiothorac Imaging. 2023 Apr 27;5(2):e220318. doi: 10.1148/ryct.220318. eCollection 2023 Apr. Radiol Cardiothorac Imaging. 2023. PMID: 37124634 Free PMC article.
-
Dual-energy CT-based stopping power prediction for dental materials in particle therapy.J Appl Clin Med Phys. 2023 Aug;24(8):e13977. doi: 10.1002/acm2.13977. Epub 2023 Apr 9. J Appl Clin Med Phys. 2023. PMID: 37032540 Free PMC article.
References
REFERENCES
-
- Barmeir E, Dubowitz B, Roffman M. Computed tomography in the assessment and planning of complicated total hip replacement. Acta Orthop. 1982;53:597-604.
-
- Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology. 1987;164:576-577.
-
- De Man B, Nuyts J, Dupont P, Marchal G, Suetens P. Metal streak artifacts in x-ray computed tomography: a simulation study. IEEE Nucl Sci Symp Med Imaging Conf. 1999;3:1860-1865.
-
- Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012;4:229-240.
-
- Kilby W, Sage J, Rabett V. Tolerance levels for quality assurance of electron density values generated from CT in radiotherapy treatment planning. Phys Med Biol. 2002;47:1485.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous