A fluorimetric assay for angiotensin-I converting enzyme in human serum
- PMID: 36243
- DOI: 10.1016/0009-8981(79)90091-3
A fluorimetric assay for angiotensin-I converting enzyme in human serum
Abstract
A fluorimetric method is described for a simple, sensitive and reproducible assay for angiotensin-I converting enzyme in human and guinea pig sera. The very weak fluorescent substrate o-aminobenzoylglycyl-p-nitro-L-phenylalanyl-L-proline is enzymatically hydrolyzed, producing the highly fluorescent o-aminobenzoylglycine that is quantitatively determined by spectrofluorimetry. Dependence of activity on substrate concentration, amount of serum, time of incubation and pH were investigated. The KM value for the substrate is 0.1 and 0.032 mM for the human and guinea pig serum enzyme, respectively. The mean value of serum angiotensin-I converting enzyme for 16 normal adult persons was 2.56 +/- 0.10 (S.E.) with a standard deviation of 0.81 nmol/min/ml serum.
Similar articles
-
A sensitive fluorimetric assay for serum angiotensin-converting enzyme.Am J Clin Pathol. 1976 Aug;66(2):416-24. doi: 10.1093/ajcp/66.2.416. Am J Clin Pathol. 1976. PMID: 181981
-
Sensitive fluorimetric assay for serum angiotensin-converting enzyme with the natural substrate angiotensin I.Am J Clin Pathol. 1977 Aug;68(2):225-8. doi: 10.1093/ajcp/68.2.225. Am J Clin Pathol. 1977. PMID: 195458
-
A fluorogenic substrate for angiotensin-converting enzyme in plasma.Clin Chem. 1978 Sep;24(9):1539-42. Clin Chem. 1978. PMID: 210982
-
A fluorescence-based protocol for quantifying angiotensin-converting enzyme activity.Nat Protoc. 2006;1(5):2423-7. doi: 10.1038/nprot.2006.349. Nat Protoc. 2006. PMID: 17406486
-
Angiotensin-converting enzyme: clinical applications and laboratory investigations on serum and other biological fluids.Crit Rev Clin Lab Sci. 1991;28(5-6):337-56. doi: 10.3109/10408369109106868. Crit Rev Clin Lab Sci. 1991. PMID: 1663362 Review.
Cited by
-
Angiotensin-converting enzyme and its clinical significance--a review.J Clin Pathol. 1983 Aug;36(8):938-47. doi: 10.1136/jcp.36.8.938. J Clin Pathol. 1983. PMID: 6308066 Free PMC article. Review.
-
Rapid screening and identification of ACE inhibitors in snake venoms using at-line nanofractionation LC-MS.Anal Bioanal Chem. 2017 Oct;409(25):5987-5997. doi: 10.1007/s00216-017-0531-3. Epub 2017 Aug 11. Anal Bioanal Chem. 2017. PMID: 28801827 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources