The influence of patient positioning on radiation dose in CT imaging: A narrative review
- PMID: 36280573
- DOI: 10.1016/j.jmir.2022.09.027
The influence of patient positioning on radiation dose in CT imaging: A narrative review
Abstract
Background and purpose: Although it is fundamental for optimal scanner operation, it is generally accepted that accurate patient centring cannot always be achieved. This review aimed to examine the reported knowledge of the negative impact of patient positioning on radiation dose and image quality during CT imaging. Furthermore, the study evaluated the current optimisation tools and techniques used to improve patient positioning relative to the gantry iso-center.
Methodology: A comprehensive search through the databases PubMed, Ovid, and Google Scholar was performed. Keywords included patient off-centring, patient positioning, localiser radiograph orientation, radiation dose, and automatic patient positioning (including synonyms). The search was limited to full-text articles that were written in English. After initial title and abstract screening, a total of 52 articles were identified to address the aim of the review. No limitations were imposed on the year of publication.
Results: Vertical off-centring was reported in up to 95% of patients undergoing chest and abdominal CT examinations, showing a significant influence on radiation dose. Depending on the scanner model and vendor, localiser orientation, bowtie filter used, and patient size, radiation dose varied from a decrease of 36% to an increase of 91%. A significant dose reduction was demonstrated when utilising an AP localiser, aligning with the trend for radiographers to off-center patients below the gantry iso-centre. Utilizing a 3D camera for body contour detection allowed for more accurate patient positioning and promoted further dose reduction.
Conclusion: Patient positioning has shown significant effects on radiation dose and image quality in CT. Developing a good understanding of the key factors influencing patient dose (off-centring direction, localiser orientation, patient size and bowtie filter selection) is critical in optimising CT scanning practices. Utilising a 3D camera for body contour detection is strongly recommended to improve patient positioning accuracy, image quality and to minimise patient dose.
Copyright © 2022. Published by Elsevier Inc.
Similar articles
-
0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.J Med Radiat Sci. 2022 Mar;69(1):5-12. doi: 10.1002/jmrs.535. Epub 2021 Aug 17. J Med Radiat Sci. 2022. PMID: 34402591 Free PMC article.
-
Localiser radiographs in CT: Current practice, radiation dose, image quality and clinical applications.Radiography (Lond). 2024 Oct;30(6):1546-1555. doi: 10.1016/j.radi.2024.09.059. Epub 2024 Oct 3. Radiography (Lond). 2024. PMID: 39366144 Review.
-
The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study.Radiography (Lond). 2024 Oct;30(6):1517-1523. doi: 10.1016/j.radi.2024.09.051. Epub 2024 Sep 19. Radiography (Lond). 2024. PMID: 39303328
-
Precise and Automatic Patient Positioning in Computed Tomography: Avatar Modeling of the Patient Surface Using a 3-Dimensional Camera.Invest Radiol. 2018 Nov;53(11):641-646. doi: 10.1097/RLI.0000000000000482. Invest Radiol. 2018. PMID: 29762259
-
The reliability of CT numbers as absolute values for diagnostic scanning, dental imaging, and radiation therapy simulation: A narrative review.J Med Imaging Radiat Sci. 2022 Mar;53(1):138-146. doi: 10.1016/j.jmir.2021.11.007. Epub 2021 Dec 13. J Med Imaging Radiat Sci. 2022. PMID: 34911666 Review.
Cited by
-
Automatic patient centering in computed tomography: a systematic review and meta-analysis.Eur Radiol. 2025 Jun;35(6):3486-3498. doi: 10.1007/s00330-024-11170-z. Epub 2024 Nov 21. Eur Radiol. 2025. PMID: 39570368
-
AI-based automatic patient positioning in a digital-BGO PET/CT scanner: efficacy and impact.EJNMMI Phys. 2025 Jan 20;12(1):4. doi: 10.1186/s40658-025-00715-w. EJNMMI Phys. 2025. PMID: 39831942 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical