Morphofunctional reaction of leukocytes and platelets in in vitro contact with a-C:H:SiOx -coated Ti-6Al-4V substrate
- PMID: 36349977
- DOI: 10.1002/jbm.a.37470
Morphofunctional reaction of leukocytes and platelets in in vitro contact with a-C:H:SiOx -coated Ti-6Al-4V substrate
Abstract
The article deals with the plasma-assisted chemical vapor deposition of 0.3-1.4 μm thick a-C:H:SiOx films in a mixture of argon and polyphenylmethylsiloxane vapor onto the Ti-6Al-4V alloy substrate, which is often used as an implant material. The a-C:H:SiOx film structure is studied by the Fourier-transform infrared and Raman spectroscopies. The pull-off adhesion test assesses the adhesive strength of a-C:H:SiOx films, and the ball-on-disk method is employed to measure their wear rate and friction coefficient. According to these studies, a-C:H:SiOx films are highly adhesive to the Ti-6Al-4V substrate, have low (0.056) friction coefficient and wear rate (9.8 × 10-8 mm3 N-1 m-1 ) in phosphate-buffered saline at 40°C. In vitro studies show neither thrombogenicity nor cytotoxicity of the a-C:H:SiOx film for the human blood mononuclear cells (hBMNCs). The in vitro contact between the hBMNC culture and a-C:H:SiOx films 0.8-1.4 μm thick deposited onto Ti-6Al-4V substrates reduces a 24-hour secretion of pro-inflammatory cytokines and chemokines IL-8, IL-17, TNFα, RANTES, and MCP-1. This reduction is more significant when the film thickness is 1.4 μm and implies its potential anti-inflammatory effect and possible application in cardiovascular surgery. The dependence is suggested for the concentration of anti-inflammatory cytokines and chemokines and the a-C:H:SiOx film thickness, which correlates with the surface wettability and electrostatic potential. The article discusses the possible applications of the anti-inflammatory effect and low thrombogenicity of a-C:H:SiOx films in cardiovascular surgery.
Keywords: a-C:H:SiOx film; biocompatibility; human blood mononuclear cells; in vitro cytotoxicity; secretory activity; thrombogenicity.
© 2022 Wiley Periodicals LLC.
References
REFERENCES
-
- Gepreel M, Niinomi M. Biocompatibility of Ti-alloys for long-term implantation. J Mech Behav Biomed Mater. 2013;20:407-415.
-
- Tian XJ, Zhang SQ, Li A, Wang HM. Effect of annealing temperature on the notch impact toughness of a laser melting deposited titanium alloy Ti-4Al-1.5Mn. Mater Sci Eng A. 2010;527:1821-1827.
-
- Kim JG, Lee KR, Yang SJ. Wear-corrosion performance of Si-DLC coatings on Ti-6Al-4V substrate. J Biomed Mater Res A. 2008;86(1):41-47.
-
- Ding HH, Fridrici V, Guillonneau G, et al. Investigation on mechanical properties of tribofilm formed on Ti-6Al-4V surface sliding against a DLC coating by nano-indentation and micro-pillar compression techniques. Wear. 2019;432-433:202954.
-
- Ding HH, Fridrici V, Geringer J, Fontaine J, Kapsa P. Influence of diamond-like carbon coatings and roughness on fretting behaviors of Ti-6Al-4V for neck adapter-femoral stem contact. Wear. 2018;406-407:53-67.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous