Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Mar;29(3):392.e1-392.e5.
doi: 10.1016/j.cmi.2022.11.003. Epub 2022 Nov 12.

Real-life evaluation of a rapid antigen test (DPP SARS-CoV-2 Antigen) for COVID-19 diagnosis of primary healthcare patients, in the context of the Omicron-dominant wave in Brazil

Affiliations

Real-life evaluation of a rapid antigen test (DPP SARS-CoV-2 Antigen) for COVID-19 diagnosis of primary healthcare patients, in the context of the Omicron-dominant wave in Brazil

Matheus Filgueira Bezerra et al. Clin Microbiol Infect. 2023 Mar.

Abstract

Objectives: We aimed to investigate the real-life performance of the rapid antigen test in the context of a primary healthcare setting, including symptomatic and asymptomatic individuals that sought diagnosis during an Omicron infection wave.

Methods: We prospectively accessed the performance of the DPP SARS-CoV-2 Antigen test in the context of an Omicron-dominant real-life setting. We evaluated 347 unselected individuals (all-comers) from a public testing centre in Brazil, performing the rapid antigen test diagnosis at point-of-care with fresh samples. The combinatory result from two distinct real-time quantitative PCR (RT-qPCR) methods was employed as a reference and 13 samples with discordant PCR results were excluded.

Results: The assessment of the rapid test in 67 PCR-positive and 265 negative samples revealed an overall sensitivity of 80.5% (CI 95% = 69.1%-89.2%), specificity of 99.2% (CI 95% = 97.3%-99.1%) and positive/negative predictive values higher than 95%. However, we observed that the sensitivity was dependent on the viral load (sensitivity in Ct < 31 = 93.7%, CI = 82.8%-98.7%; Ct > 31 = 47.4%, CI = 24.4%-71.1%). The positive samples evaluated in the study were Omicron (BA.1/BA.1.1) by whole-genome sequencing (n = 40) and multiplex RT-qPCR (n = 17).

Conclusions: Altogether, the data obtained from a real-life prospective cohort supports that the rapid antigen test sensitivity for Omicron remains high and underscores the reliability of the test for COVID-19 diagnosis in settings with high disease prevalence and limited PCR testing capability.

Keywords: COVID-19 diagnosis; Evidence-based; Omicron; Rapid antigen test; Viral control.

PubMed Disclaimer

Figures

Image 1
Graphical abstract
Fig. 1
Fig. 1
Study design and DPP SARS-CoV-2 Antigen performance rates. (A) The organogram displays the number of nasopharyngeal samples evaluated and their distribution according to inclusion criteria, diagnostic test results and sequencing. Fifteen percent of the positive samples (n = 10) could not be assessed for the SARS-CoV-2 variant due to the low viral load (B) Performance rates observed in the study. The bars represent the 95% confidence intervals.
Fig. 2
Fig. 2
Comparison of the PCR and the rapid antigen test results. (A) Correlation between the Ct values from the two PCR methods that were used combined as the reference method. The colourless datapoint represents the outlier sample (B) Distribution of Ct values and the proportion of positive samples in the rapid test. The N1 target (CDC protocol) is more skewed towards lower Ct values whereas the E target (EP/RP kit) is more skewed towards higher Ct Values. The dashed line represents the average values of the Ct values from positive samples. (C) Ct values grouped according to the rapid test result: Ag-positive samples had lower Ct values (higher viral load) than the Ag-negative samples. (D) Probability of positive Ag test according to the Ct values (N1 target) in groups of 3 (e.g. 16–18, 19–21). The curve was calculated by the non-linear asymmetric sigmoidal curve fit.
Figure S1
Figure S1
Figure S2
Figure S2
Figure S3
Figure S3

References

    1. Koo J.R., Cook A.R., Lim J.T., Tan K.W., Dickens B.L. Modelling the impact of mass testing to transition from pandemic mitigation to endemic COVID-19. Viruses. 2022;14:967. doi: 10.3390/v14050967. - DOI - PMC - PubMed
    1. Raïch-Regué D., Muñoz-Basagoiti J., Perez-Zsolt D., Noguera-Julian M., Pradenas E., Munoz E.R., et al. Performance of SARS-CoV-2 antigen-detecting rapid diagnostic tests for omicron and other variants of concern. Front Microbiol. 2022;13 doi: 10.3389/fmicb.2022.810576. - DOI - PMC - PubMed
    1. Wang L., Cheng G. Sequence analysis of the emerging SARS-CoV-2 variant Omicron in South Africa. J Med Virol. 2022;94:1728–1733. doi: 10.1002/jmv.27516. - DOI - PubMed
    1. World Health Organization . 2021. Antigen-detection in the diagnosis of SARS-CoV-2 infection: interim guidance.https://apps.who.int/iris/handle/10665/345948 Accessed March 20, 2022.
    1. European Commission Directorate-General For Health And Food Safety . 2022. EU Common list of COVID-19 rapid antigen tests and a list of mutually recognised COVID-19 laboratory based antigenic assays.https://health.ec.europa.eu/health-security-and-infectious-diseases/cris... Updated June 10.

LinkOut - more resources