A 3D stack-of-spirals approach for rapid hyperpolarized 129 Xe ventilation mapping in pediatric cystic fibrosis lung disease
- PMID: 36433705
- DOI: 10.1002/mrm.29505
A 3D stack-of-spirals approach for rapid hyperpolarized 129 Xe ventilation mapping in pediatric cystic fibrosis lung disease
Abstract
Purpose: To demonstrate the feasibility of a rapid 3D stack-of-spirals (3D-SoS) imaging acquisition for hyperpolarized 129 Xe ventilation mapping in healthy pediatric participants and pediatric cystic fibrosis (CF) participants, in comparison to conventional Cartesian multislice (2D) gradient-recalled echo (GRE) imaging.
Methods: The 2D-GRE and 3D-SoS acquisitions were performed in 13 pediatric participants (5 healthy, 8 CF) during separate breath-holds. Images from both sequences were compared on the basis of ventilation defect percent (VDP) and other measures of image similarity. The nadir of transient oxygen saturation (SpO2 ) decline due to xenon breath-holding was measured with pulse oximetry, and expressed as a percent change relative to baseline.
Results: 129 Xe ventilation images were acquired in a breath-hold of 1.2-1.8 s with the 3D-SoS sequence, compared to 6.2-8.8 s for 2D-GRE. Mean ± SD VDP measures for 2D-GRE and 3D-SoS sequences were 5.02 ± 1.06% and 5.28 ± 1.08% in healthy participants, and 18.05 ± 8.26% and 18.75 ± 6.74% in CF participants, respectively. Across all participants, the intraclass correlation coefficient of VDP measures for both sequences was 0.98 (95% confidence interval: 0.94-0.99). The percent change in SpO2 was reduced to -2.1 ± 2.7% from -5.2 ± 3.5% with the shorter 3D-SoS breath-hold.
Conclusion: Hyperpolarized 129 Xe ventilation imaging with 3D-SoS yielded images approximately five times faster than conventional 2D-GRE, reducing SpO2 desaturation and improving tolerability of the xenon administration. Analysis of VDP and other measures of image similarity demonstrate excellent agreement between images obtained with both sequences. 3D-SoS holds significant potential for reducing the acquisition time of hyperpolarized 129 Xe MRI, and/or increasing spatial resolution while adhering to clinical breath-hold constraints.
Keywords: cystic fibrosis; hyperpolarized 129Xe; pediatric; spiral; ventilation defect.
© 2022 International Society for Magnetic Resonance in Medicine.
References
REFERENCES
-
- Fain SB, Korosec FR, Holmes JH, O'Halloran R, Sorkness RL, Grist TM. Functional lung imaging using hyperpolarized gas MRI. J Magn Reson Imaging. 2007;25:910-923. doi:10.1002/jmri.20876
-
- Mugler JP III, Altes TA. Hyperpolarized 129Xe MRI of the human lung. J Magn Reson Imaging. 2013;37:313-331. doi:10.1002/jmri.23844
-
- Kruger SJ, Nagle SK, Couch MJ, Ohno Y, Albert M, Fain SB. Functional imaging of the lungs with gas agents. J Magn Reson Imaging. 2016;43:295-315. doi:10.1002/jmri.25002
-
- Ebner L, Kammerman J, Driehuys B, Schiebler ML, Cadman RV, Fain SB. The role of hyperpolarized 129 xenon in MR imaging of pulmonary function. Eur J Radiol. 2017;86:343-352. doi:10.1016/j.ejrad.2016.09.015
-
- Walkup LL, Thomen RP, Akinyi TG, et al. Feasibility, tolerability and safety of pediatric hyperpolarized 129Xe magnetic resonance imaging in healthy volunteers and children with cystic fibrosis. Pediatr Radiol. 2016;46:1651-1662. doi:10.1007/s00247-016-3672-1
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous