CRT-D replacement strategy: results of the BioCONTINUE study
- PMID: 36459310
- DOI: 10.1007/s10840-022-01440-5
CRT-D replacement strategy: results of the BioCONTINUE study
Abstract
Background: In patients with cardiac resynchronization therapy defibrillators (CRT-Ds), the need for implantable cardioverter-defibrillator (ICD) back-up may be questionable at time of CRT-D replacement (REP) if ICD implant criteria are no longer met due to an improved left ventricular ejection fraction (LVEF) and if no major ventricular arrhythmic event (VAE) occurred during the CRT-D lifetime. The aim of our study was to assess the relevance of ICD back-up and predictors of VAE after REP in primary prevention CRT-D patients.
Methods: The prospective, observational, international BioCONTINUE study investigated the rate of patients with at least 1 sustained VAE (sVAE) post-REP and searched for predictive factors of sVAE.
Results: Two hundred seventy-six patients (70 ± 10 years, 77% men, mean LVEF 40.6 ± 12.6%) were followed for 28.4 ± 10.2 months. The rate of patients with sVAE was 8.3%, 10.3%, and 21.2% at 1, 2, and 4 years post-REP. Patients without persistent ICD indication at REP still had a sVAE rate of 5.7% (95% CI 2.3-11.5%) at 2 years. In multivariate analysis, predictive factors of subsequent sVAE were (i) persistent ICD indication (hazard ratio (HR) 3.6; 95% CI 1.6-8.3; p = 0.003); (ii) 64-72 years of age as compared to ≥ 79 years (HR 3.7; 95% CI 1.4-9.7; p = 0.008); and (iii) ischemic heart disease (HR 4.4; 95% CI 2.1-9.3; p < 0.0001).
Conclusions: The risk of sVAE (21.2% at 4 years post-REP) depends on age, ischemic heart disease, and ICD indication at the time of REP. A non-trivial risk of sVAE remains in patients without persistent ICD indication.
Clinical trial registration: NCT02323503.
Keywords: Cardiac resynchronization therapy (CRT); ICD replacement; Implantable cardioverter-defibrillator (ICD); Primary ICD indications; Sudden death; Ventricular arrhythmia.
© 2022. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.
Comment in
-
Downgrade from CRT-D to CRT-P at time of generator replacement: a true clinical conundrum or just an academic debate?J Interv Card Electrophysiol. 2023 Sep;66(6):1315-1316. doi: 10.1007/s10840-023-01506-y. Epub 2023 Feb 16. J Interv Card Electrophysiol. 2023. PMID: 36795270 No abstract available.
References
-
- Manfredi JA, Al-Khatib SM, Shaw LK, Thomas L, Fogel RI, Padanilam B, et al. Association between left ventricular ejection fraction post-cardiac resynchronization treatment and subsequent implantable cardioverter defibrillator therapy for sustained ventricular tachyarrhythmias. Circ Arrhythm Electrophysiol. 2013;6:257–64. - DOI - PubMed
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous