Error detection using EPID-based 3D in vivo dose verification for lung stereotactic body radiotherapy

Affiliations

Error detection using EPID-based 3D in vivo dose verification for lung stereotactic body radiotherapy

Jianghua Huang et al. Appl Radiat Isot. 2023 Feb.

Abstract

Purpose: To investigate the error detectability limitations of an EPID-based 3D in vivo dosimetry verification system for lung stereotactic body radiation therapy (SBRT).

Methods: Thirty errors were intentionally introduced, consisting of dynamic and constant machine errors, to simulate the possible errors that may occur during delivery. The dynamic errors included errors in the output, gantry angle and MLC positions related to gantry inertial and gravitational effects, while the constant errors included errors in the collimator angle, jaw positions, central leaf positions, setup shift and thickness to simulate patient weight loss. These error plans were delivered to a CIRS phantom using the SBRT technique for lung cancer. Following irradiation of these error plans, the dose distribution was reconstructed using iViewDose™ and compared with the no error plan.

Results: All errors caused by the central leaf positions, dynamic MLC errors, Jaw inwards movements, setup shifts and patient anatomical changes were successfully detected. However, dynamic gantry angle and collimator angle errors were not detected in the lung case due to the rotation-symmetric target shape. The results showed that the γmean and γpassrate indicators can detect 13 (81.3%) and 14 (87.5%) of the 16 errors respectively without including the gantry angle error, collimator angle error and output error.

Conclusions: In summary, iViewDose™ is an appropriate approach for detecting most types of clinical errors for lung SBRT. However, the phantom results also showed some detectability limitations of the system in terms of dynamic gantry angle and constant collimator angle errors.

Keywords: EPID; Error detection; In vivo dosimetry; Lung SBRT; Quality assurance.