Error detection using EPID-based 3D in vivo dose verification for lung stereotactic body radiotherapy
- PMID: 36459899
- DOI: 10.1016/j.apradiso.2022.110567
Error detection using EPID-based 3D in vivo dose verification for lung stereotactic body radiotherapy
Abstract
Purpose: To investigate the error detectability limitations of an EPID-based 3D in vivo dosimetry verification system for lung stereotactic body radiation therapy (SBRT).
Methods: Thirty errors were intentionally introduced, consisting of dynamic and constant machine errors, to simulate the possible errors that may occur during delivery. The dynamic errors included errors in the output, gantry angle and MLC positions related to gantry inertial and gravitational effects, while the constant errors included errors in the collimator angle, jaw positions, central leaf positions, setup shift and thickness to simulate patient weight loss. These error plans were delivered to a CIRS phantom using the SBRT technique for lung cancer. Following irradiation of these error plans, the dose distribution was reconstructed using iViewDose™ and compared with the no error plan.
Results: All errors caused by the central leaf positions, dynamic MLC errors, Jaw inwards movements, setup shifts and patient anatomical changes were successfully detected. However, dynamic gantry angle and collimator angle errors were not detected in the lung case due to the rotation-symmetric target shape. The results showed that the γmean and γpassrate indicators can detect 13 (81.3%) and 14 (87.5%) of the 16 errors respectively without including the gantry angle error, collimator angle error and output error.
Conclusions: In summary, iViewDose™ is an appropriate approach for detecting most types of clinical errors for lung SBRT. However, the phantom results also showed some detectability limitations of the system in terms of dynamic gantry angle and constant collimator angle errors.
Keywords: EPID; Error detection; In vivo dosimetry; Lung SBRT; Quality assurance.
Copyright © 2022 Elsevier Ltd. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Similar articles
-
EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans.Phys Med. 2019 Mar;59:37-46. doi: 10.1016/j.ejmp.2019.02.007. Epub 2019 Feb 26. Phys Med. 2019. PMID: 30928064
-
A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry : A step to QA with in vivo EPID dosimetry.Strahlenther Onkol. 2020 Feb;196(2):182-192. doi: 10.1007/s00066-019-01549-z. Epub 2020 Jan 10. Strahlenther Onkol. 2020. PMID: 31925465
-
Feasibility study of an electronic portal imaging based in vivo dose verification system for prostate stereotactic body radiotherapy.Phys Med. 2019 Aug;64:204-209. doi: 10.1016/j.ejmp.2019.07.008. Epub 2019 Jul 24. Phys Med. 2019. PMID: 31515021
-
Error detection during VMAT delivery using EPID-based 3D transit dosimetry.Phys Med. 2018 Oct;54:137-145. doi: 10.1016/j.ejmp.2018.10.005. Epub 2018 Oct 11. Phys Med. 2018. PMID: 30337003
-
Estimating dose delivery accuracy in stereotactic body radiation therapy: A review of in-vivo measurement methods.Radiother Oncol. 2020 Aug;149:158-167. doi: 10.1016/j.radonc.2020.05.014. Epub 2020 May 19. Radiother Oncol. 2020. PMID: 32416282 Review.
Cited by
-
Does deformation of immobilization devices impact treatment accuracy in thoracic cancer radiotherapy?J Appl Clin Med Phys. 2025 Apr;26(4):e14619. doi: 10.1002/acm2.14619. Epub 2024 Dec 18. J Appl Clin Med Phys. 2025. PMID: 39693569 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical