. 2022 Dec 12;10(1):52.
doi: 10.1186/s40635-022-00479-y.

Effects of changes in inspired oxygen fraction on urinary oxygen tension measurements

Affiliations

Effects of changes in inspired oxygen fraction on urinary oxygen tension measurements

Eduardo A Osawa et al. Intensive Care Med Exp. .

Abstract

Background: Continuous measurement of urinary PO2 (PuO2) is being applied to indirectly monitor renal medullary PO2. However, when applied to critically ill patients with shock, its measurement may be affected by changes in FiO2 and PaO2 and potential associated O2 diffusion between urine and ureteric or bladder tissue. We aimed to investigate PuO2 measurements in septic shock patients with a fiberoptic luminescence optode inserted into the urinary catheter lumen in relation to episodes of FiO2 change. We also evaluated medullary and urinary oxygen tension values in Merino ewes at two different FiO2 levels.

Results: In 10 human patients, there were 32 FiO2 decreases and 31 increases in FiO2. Median pre-decrease FiO2 was 0.36 [0.30, 0.39] and median post-decrease FiO2 was 0.30 [0.23, 0.30], p = 0.006. PaO2 levels decreased from 83 mmHg [77, 94] to 72 [62, 80] mmHg, p = 0.009. However, PuO2 was 23.2 mmHg [20.5, 29.0] before and 24.2 mmHg [20.6, 26.3] after the intervention (p = 0.56). The median pre-increase FiO2 was 0.30 [0.21, 0.30] and median post-increase FiO2 was 0.35 [0.30, 0.40], p = 0.008. PaO2 levels increased from 64 mmHg [58, 72 mmHg] to 71 mmHg [70, 100], p = 0.04. However, PuO2 was 25.0 mmHg [IQR: 20.7, 26.8] before and 24.3 mmHg [IQR: 20.7, 26.3] after the intervention (p = 0.65). A mixed linear regression model showed a weak correlation between the variation in PaO2 and the variation in PuO2 values. In 9 Merino ewes, when comparing oxygen tension levels between FiO2 of 0.21 and 0.40, medullary values did not differ (25.1 ± 13.4 mmHg vs. 27.9 ± 15.4 mmHg, respectively, p = 0.6766) and this was similar to urinary oxygen values (27.1 ± 6.17 mmHg vs. 29.7 ± 4.41 mmHg, respectively, p = 0.3192).

Conclusions: Changes in FiO2 and PaO2 within the context of usual care did not affect PuO2. Our findings were supported by experimental data and suggest that PuO2 can be used as biomarker of medullary oxygenation irrespective of FiO2.

Keywords: Inspired oxygen fraction; Sepsis; Septic shock; Urinary oxygen tension; Urinary oxygenation.

PubMed Disclaimer

Conflict of interest statement

The authors have no conflicts of interest to declare.

Figures

Fig. 1
Fig. 1
Schematic representation of the procedure used to obtain PaO2 and PuO2 measurements before and after an episode of FiO2 change during the observation period. ABG: arterial blood gas; FiO2: fraction of inspired oxygen; PuO2: urinary oxygen tension
Fig. 2
Fig. 2
Scatterplot of arterial and urinary oxygen tension measured in the ten patients enrolled in the study. PuO2: urinary oxygen tension; PaO2: arterial blood oxygen tension
Fig. 3
Fig. 3
Box plot representation of arterial blood oxygen tension before and after changes in fractional inspired oxygen. FiO2: fractional inspired oxygen; PaO2: arterial blood oxygen tension
Fig. 4
Fig. 4
Box plot representation of urinary oxygen tension before and after changes in fractional inspired oxygen. FiO2: fractional inspired oxygen; PuO2: urinary oxygen tension
Fig. 5
Fig. 5
Relationship between the changes in arterial oxygen tension (Δ PaO2) and urinary oxygen tension (Δ PuO2)

References

    1. Schrier RW, Wang W. Acute renal failure and sepsis. N Engl J Med. 2004;351(2):159–169. doi: 10.1056/NEJMra032401. - DOI - PubMed
    1. Ow CPC, Ngo JP, Ullah MM, Hilliard LM, Evans RG. Renal hypoxia in kidney disease: cause or consequence? Acta Physiol. 2018;222(4):e12999. doi: 10.1111/apha.12999. - DOI - PubMed
    1. Calzavacca P, Evans RG, Bailey M, Bellomo R, May CN. Cortical and medullary tissue perfusion and oxygenation in experimental septic acute kidney injury. Crit Care Med. 2015;43(10):e431–439. doi: 10.1097/CCM.0000000000001198. - DOI - PubMed
    1. Lankadeva YR, Kosaka J, Evans RG, Bellomo R, May CN. Urinary oxygenation as a surrogate measure of medullary oxygenation during angiotensin II therapy in septic acute kidney injury. Crit Care Med. 2018;46(1):e41–e48. doi: 10.1097/CCM.0000000000002797. - DOI - PubMed
    1. Zhu MZL, Martin A, Cochrane AD, et al. Urinary hypoxia: an intraoperative marker of risk of cardiac surgery-associated acute kidney injury. Nephrol Dial Transplant. 2018;33(12):2191–2201. doi: 10.1093/ndt/gfy047. - DOI - PubMed

LinkOut - more resources