. 2023 Jan 25:1689:463756.
doi: 10.1016/j.chroma.2022.463756. Epub 2022 Dec 29.

Performance of global retention models in the optimisation of the chromatographic separation (I): Simple multi-analyte samples

Affiliations
Free article

Performance of global retention models in the optimisation of the chromatographic separation (I): Simple multi-analyte samples

P Peiró-Vila et al. J Chromatogr A. .
Free article

Abstract

Conventional retention models lead to accurate descriptions of the elution behaviour from the fitting of data for single solutes or from a set of solutes, one by one. However, the simultaneous fitting of several solutes through a regression process that separates the contributions of column and solvent from those of each solute is also possible. The result is a global retention model constituted by a set of equations with some common parameters (those associated with column and solvent), whereas others, specific to each solute, differ for each equation. This work explores the possibilities, advantages, and limitations of global models when they are applied to the optimisation of chromatographic resolution. A set constituted by 13 drugs (diuretics and β-blockers) and a training experimental design of seven multi-linear gradients are considered. Since standards for all compounds were available, the optimisation based on global models could be compared with the conventional optimisation, which is based on individual models. In their current state, global models do not predict changes in elution order, but they do allow for incorporating additional solutes (e.g., new analytes or matrix peaks) with only one new experiment. This possibility is explored by extending the model for the 13 analytes to include 26 peaks associated with a contamination in the injector. The combination of individual and global models allows an optimisation where the effects of matrix peaks on the separation of analytes can be integrated.

Keywords: Global retention models; Modelling of unknowns; Multi-analyte samples; Multi-linear gradient elution; Optimisation of resolution.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.