High precision but systematic offset in a standing bioelectrical impedance analysis (BIA) compared with dual-energy X-ray absorptiometry (DXA)
- PMID: 36619314
- PMCID: PMC9813632
- DOI: 10.1136/bmjnph-2022-000512
High precision but systematic offset in a standing bioelectrical impedance analysis (BIA) compared with dual-energy X-ray absorptiometry (DXA)
Abstract
Bioelectrical impedance analysis (BIA) provides a practical method of body composition estimation for field research and weight management programmes, with devices and algorithms that have improved in recent years. We compared suitability of a commercial BIA system that uses multi-frequency-based proprietary algorithms (InBody 770, Cerritos, California, USA) and a laboratory-based validated single-frequency system (Quantum IV, RJL Systems, Clinton Township, Michigan, USA) with dual-energy X-ray absorptiometry (DXA) (iDXA, GE Lunar, Madison, Wisconsin, USA). Volunteers included fit non-obese active duty US Marines (480 men; 315 women), assessed by DXA and the two BIA systems. Both RJL and InBody BIA devices predicted DXA-based fat-free mass (FFM) (mean absolute error (MAE) 2.8 and 3.1 kg, respectively) and per cent body fat (%BF) (MAE 3.4% and 3.9%, respectively), with higher correlations from the InBody device (r2=0.96 (%BF) and 0.84 (FFM)) versus the RJL (r2=0.92 (%BF) and 0.72 (FFM)). InBody overpredicted FFM (bias +2.7, MAE 3.1 kg) and underpredicted %BF (bias -3.4 and MAE 3.9%) versus the RJL. A 3% correction factor applied to the InBody device results provided values very close to the DXA measurements. These findings support the application of modern BIA systems to body composition goals of maximum %BF and minimum lean body mass for both men and women.
Keywords: Nutrition assessment; Physical performance; Weight management.
© Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: None declared.
Figures
References
-
- Thomasset MA. [Bioelectric properties of tissue. Impedance measurement in clinical medicine. Significance of curves obtained]. Lyon Med 1962;94:107–18. - PubMed
-
- Nyboer J. Electrical impedance plethysmography. In: Thomas CC, ed. Springfield, ill, 1959.
-
- Moore FD. The body cell mass and its supporting environment. London: W.B. Saunders, 1963.
-
- Mills W, Rau D. University of Alaska high latitude study, and the MT. McKinley project, 1981-1983. Alaska Med 1983;25:21–8. - PubMed
LinkOut - more resources
Full Text Sources
Miscellaneous