An evaluation of six disposable heat and moisture exchangers
- PMID: 3661965
- DOI: 10.1177/0310057X8701500312
An evaluation of six disposable heat and moisture exchangers
Abstract
Six disposable heat and moisture exchangers were tested on patients undergoing anaesthesia requiring mechanical ventilation. Inspiratory humidity and temperature were monitored to find the steady-state values reached with each device together with the times taken for these to be achieved. The exchangers were tested in a non-rebreathing T-piece circuit and in a conventional circle system with a fresh gas flow of 6 l/min: the Siemens 150 provided 28 and 32 mg of water/litre of inspired gas (at about 30 degrees C) in 10 and 5 min respectively, but is rather heavy and bulky. The Portex Humidvent provided 25 and 30 mg/l, and although taking longer to reach steady state (27 and 15 min respectively) is small, light and cheap. The Siemens 151 provided 25 and 27 mg/l in 18 and 10 min respectively, but is heavier than the Portex exchanger. The performances of these three devices were not significantly different from each other in either study (P less than 0.05). For the T-piece system the Pall and Engstrom exchangers were the next most efficient. The Pall device provided 18 and 23 mg/l (in 18 and 8 min respectively) and the Engstrom provided 20 and 23 mg/l (in 19 and 10 min respectively). In the circle system, there were no significant differences between the performances of the Portex, Siemens 151, Pall and Engstrom exchangers. The Pall is also a very effective bacterial filter and has been found to be satisfactory in the intensive care setting. The Terumo appeared to perform no better than a circle system with catheter mount (13 mg/l at 27 degrees C). It would seem that more complex humidification equipment is not necessary during anaesthesia if an efficient heat and moisture exchanger is used.
Similar articles
-
Preservation of humidity and heat of respiratory gases in patients with a minute ventilation greater than 10 L/min.Crit Care Med. 1994 Nov;22(11):1871-6. Crit Care Med. 1994. PMID: 7956294 Clinical Trial.
-
Comparing two heat and moisture exchangers with one vaporizing humidifier in patients with minute ventilation greater than 10 L/min.Chest. 1995 May;107(5):1411-5. doi: 10.1378/chest.107.5.1411. Chest. 1995. PMID: 7750340 Clinical Trial.
-
Efficacy of nine heat and moisture exchangers for intraoperative airway heat conservation.J Aerosol Med. 1991 Summer;4(2):117-25. doi: 10.1089/jam.1991.4.117. J Aerosol Med. 1991. PMID: 10147688
-
Humidification during anesthesia.Respir Care Clin N Am. 1998 Jun;4(2):281-99. Respir Care Clin N Am. 1998. PMID: 9648187 Review.
-
Heat and moisture exchangers and breathing system filters: their use in anaesthesia and intensive care. Part 2 - practical use, including problems, and their use with paediatric patients.Anaesthesia. 2011 Jan;66(1):40-51. doi: 10.1111/j.1365-2044.2010.06564.x. Epub 2010 Nov 30. Anaesthesia. 2011. PMID: 21118189 Review.
Cited by
-
Effects of surgical site and inspired gas warming devices on body temperature during lower abdominal and thoracic surgery.J Anesth. 1992 Oct;6(4):467-73. doi: 10.1007/s0054020060467. J Anesth. 1992. PMID: 15278521
-
Measurement of tracheal temperature is not a reliable index of total respiratory heat loss in mechanically ventilated patients.Crit Care. 2001;5(1):24-30. doi: 10.1186/cc974. Epub 2000 Dec 8. Crit Care. 2001. PMID: 11178222 Free PMC article. Clinical Trial.
-
Effects of the heat-moisture exchangers on dynamic hyperinflation of mechanically ventilated COPD patients.Intensive Care Med. 1990;16(7):441-3. doi: 10.1007/BF01711222. Intensive Care Med. 1990. PMID: 2269712
-
A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings.Intensive Care Med. 1998 Feb;24(2):138-46. doi: 10.1007/s001340050535. Intensive Care Med. 1998. PMID: 9539071
-
Use of heat and moisture exchanging (HME) filters in mechanically ventilated ICU patients: influence on airway flow-resistance.Intensive Care Med. 1993;19(8):462-6. doi: 10.1007/BF01711088. Intensive Care Med. 1993. PMID: 8294629
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources