. 2022 Nov 29;17(4):536-542.
doi: 10.18502/jovr.v17i4.12314. eCollection 2022 Oct-Dec.

Clinical Evaluation of the 3nethra Aberro Handheld Autorefractometer

Affiliations

Clinical Evaluation of the 3nethra Aberro Handheld Autorefractometer

Selvamani Perumal et al. J Ophthalmic Vis Res. .

Abstract

Purpose: To evaluate the 3nethra aberro auto refractometer device as an alternative tool for quick and reliable measurement of refractive errors and to compare it with the gold standard subjective refractive error measurement.

Methods: Refractive errors were measured using both subjective refraction and the 3nethra aberro handheld autorefractometer. The refractive measurements were converted into equivalent vector notations of spherical equivalent and Jackson cross-cylinder measurements J0 & J45. The resultant power vectors were compared with subjective measurements.

Results: This clinical study comprised 60 subjects (22 male and 38 female; with a mean age of 34 ± 16 years). Data, when compared with the subjective refraction measurements, resulted in 90% of power vectors values in both left and right eyes being the same in the 3nethra aberro handheld autorefractometer and the subjective measurement. The refractive error measurements also had an agreement of 70% and 90% when the range of diopter was between ± 0.25 and ± 0.5D, respectively. When the Bland-Altman's plot analysis was performed, about 98% of data lied within the ± 2 standard deviation variation. An average correlation between the two methods of error measurement was 0.74, and the paired t-test showed P > 0.05 for all the power vectors except for the spherical equivalent in the right eye.

Conclusion: The 90% agreement between the error measurements done by two methods indicates that the 3nethra aberro handheld autorefractometer can function as an alternative for the time-consuming subjective refractive error measurement.

Keywords: Non-mydriatic; Pupil Diameter; Refractive Index; Subjective Refractive Error; Vision; Wavefront Technology; Aberro Autorefractometer.

PubMed Disclaimer

Conflict of interest statement

None declared.

Figures

Figure 1
Figure 1
The Bland-Altman plots for comparing the SR and AHAR observations in the right and left eyes for SPE. 95% limits of the agreement are indicated by the upper and lower dashed line, and the mean is indicated by a solid line.
Figure 2
Figure 2
The correlation of the three power vectors in both the eyes.

References

    1. Resnikoff S, Pascolini D, Mariotti SP, Pokharel GP. Global magnitude of visual impairment caused by uncorrected refractive errors in 2004. Bull World Health Organ . 2008;86:63–70. - PMC - PubMed
    1. Bao J, Le R, Wu J, Shen Y, Lu F, He JC. Higher-order wavefront aberrations for populations of young emmetropes and myopes. J Optom . 2009;2:51–58.
    1. Bennett JR, Stalboerger GM, Hodge DO, Schornack MM. Comparison of refractive assessment by wavefront aberrometry, autorefraction, and subjective refraction. J Optom . 2015;8:109–115. - PMC - PubMed
    1. Lebow KA, Campbell CE. A comparison of a traditional and wavefront autorefraction. Optom Vis Sci. 2014;91:1191–1198. - PubMed
    1. Salmon TO, van de Pol C. Evaluation of a clinical aberrometer for lower-order accuracy and repeatability, higher-order repeatability, and instrument myopia. Optometry. 2005;76:461–472. - PubMed

LinkOut - more resources