Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue
- PMID: 36640426
- DOI: 10.1111/jce.15813
Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue
Abstract
Introduction: Contact force has been used to titrate lesion formation for radiofrequency ablation. Pulsed field ablation (PFA) is a field-based ablation technology for which limited evidence on the impact of contact force on lesion size is available.
Methods: Porcine hearts (n = 6) were perfused using a modified Langendorff set-up. A prototype focal PFA catheter attached to a force gauge was held perpendicular to the epicardium and lowered until contact was made. Contact force was recorded during each PFA delivery. Matured lesions were cross-sectioned, stained, and the lesion dimensions measured.
Results: A total of 82 lesions were evaluated with contact forces between 1.3 and 48.6 g. Mean lesion depth was 4.8 ± 0.9 mm (standard deviation), mean lesion width was 9.1 ± 1.3 mm, and mean lesion volume was 217.0 ± 96.6 mm3 . Linear regression curves showed an increase of only 0.01 mm in depth (depth = 0.01 × contact force + 4.41, R2 = 0.05), 0.03 mm in width (width = 0.03 × contact force + 8.26, R2 = 0.13) for each additional gram of contact force, and 2.20 mm3 in volume (volume = 2.20 × contact force + 162, R2 = 0.10).
Conclusion: Increasing contact force using a bipolar, biphasic focal PFA system has minimal effects on acute lesion dimensions in an isolated porcine heart model and achieving tissue contact is more important than the force with which that contact is made.
Keywords: cardiac ablation; contact force; irreversible electroporation; pulsed field ablation.
© 2023 Medtronic Inc and The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC.
Comment in
-
Pulsed-field ablation: A little force to be reckoned with.J Cardiovasc Electrophysiol. 2023 May;34(5):1320-1321. doi: 10.1111/jce.15864. Epub 2023 Feb 23. J Cardiovasc Electrophysiol. 2023. PMID: 36807629 No abstract available.
References
REFERENCES
-
- Haines DE. Letter regarding article, “direct measurement of the lethal isotherm for radiofrequency ablation of myocardial tissue. Circulation: Arrhythmia Electrophysiol. 2011;4:e67.
-
- Wood M, Goldberg S, Lau M, et al. Direct measurement of the lethal isotherm for radiofrequency ablation of myocardial tissue. Circulation: Arrhythmia Electrophysiol. 2011;4:373-378.
-
- Haines DE. Determinants of lesion size during radiofrequency catheter ablation: the role of electrode-tissue contact pressure and duration of energy delivery. J Cardiovasc Electrophysiol. 1991;2:509-515.
-
- Avitall B, Mughal K, Hare J, Helms R, Krum D. The effects of electrode-tissue contact on radiofrequency lesion generation. Pacing Clin Electrophysiol. 1997;20:2899-2910.
-
- Strickberger SA, Vorperian VR, Man KC, et al. Relation between impedance and endocardial contact during radiofrequency catheter ablation. Am Heart J. 1994;128:226-229.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources