Rationale for and use of the lumenless 3830 pacing lead

Affiliations
Review

Rationale for and use of the lumenless 3830 pacing lead

Travis D Richardson et al. J Cardiovasc Electrophysiol. 2023 Mar.

Abstract

Most currently available pacing and defibrillation leads utilize a stylet-based design that facilitates implantation. This has advantages, but also increases the lead diameter and adds the potential for metal fatigued-based conductor failure. A systematic literature search was conducted, and the authors add their twenty-year experience with this lead design. The global experience with lumenless leads was reviewed both for "standard" positioning and with conduction system pacing. Methods for both placement and system modification are reviewed. Lumenless leads have the potential to improve the durability of endocardial pacing and facilitate conduction system pacing.

Keywords: catheters; lead failure; pacemaker leads.

PubMed Disclaimer

References

REFERENCES

    1. Cantu F, De Filippo P, Gabbarini F, et al. Selective-site pacing in paediatric patients: a new application of the Select Secure system. Europace. 2009;11(5):601-606.
    1. Bharmanee A, Zelin K, Sanil Y, Gupta P, Karpawich PP. Comparative chronic valve and venous effects of lumenless versus stylet-delivered pacing leads in patients with and without congenital heart. Pacing Clin Electrophysiol. 2015;38(11):1343-1350.
    1. Gabbarini F, Agnoletti G. Selective-site pacing in paediatric patients: use of the SelectSecure System and risk of vein occlusion. Europace. 2010;12(9):1286-1289.
    1. Gammage MD, Lieberman RA, Yee R, et al. Multi-center clinical experience with a lumenless, catheter-delivered, bipolar, permanent pacemaker lead: implant safety and electrical performance. Pacing Clin Electrophysiol. 2006;29(8):858-865.
    1. Bansal N, Samuel S, Zelin K, Karpawich PP. Ten-year clinical experience with the lumenless, catheter-delivered, 4.1-Fr diameter pacing lead in patients with and without congenital heart. Pacing Clin Electrophysiol. 2017;40(1):17-25.