Accuracy of Four Apex Locators in Premolars with Root Resorption: An In vitro Study
- PMID: 36703871
- PMCID: PMC9868988
- DOI: 10.22037/iej.v17i1.35498
Accuracy of Four Apex Locators in Premolars with Root Resorption: An In vitro Study
Abstract
Introduction: The aim of the present in vitro study was to compare the accuracy of the working length measurements of four foramen locators: Root ZX mini, Raypex 6, Woodpex III and Propex Pixi in uniradicular premolars with simulated root resorption.
Materials and methods: For this study, 30 single-rooted permanent premolars were selected. The samples were divided into two groups: 15 teeth with simulated external root resorption and 15 teeth with internal resorption. Each sample was immersed in containers with alginate covering only the root part. Measurements were acquired from the four foramen locators and compared with the working length measurements that have been previously obtained with the direct visual technique and the use of a Digital USB Microscope under 5×. Next, the Shapiro-Wilk normality test was applied. The ANOVA test and t-test for related samples were performed, in order to analyze the measurements obtained.
Results: In the teeth with external resorption, no statistically significant differences in the measurements were obtained with Raypex 6 and Propex Pixi foramen locators, but there were statistically significant differences with Root ZX mini and Woodpex III Root Zx mini [Root Zx mini (P=0.040) and Woodpex III (P=0.000)]. On the other hand, in the samples with internal root resorption, there were no statistically significant differences in measurements with the Root ZX mini, Propex Pixi and Raypex 6 foramen locators. However, there were significant differences with Woodpex III.
Conclusions: Based on this in vitro study, Raypex 6 had the highest accuracy in premolars with simulated external resorption and Root ZX mini was the most accurate in teeth with simulated internal resorption. Furthermore, external root resorption affects the accuracy of foramen locators more than internal resorption.
Keywords: Apical Foramen; Electronic Apex Locator; Root Resorption.
Conflict of interest statement
‘None declared’.
Figures
Similar articles
-
Clinical Accuracy and Precision of 3 Multifrequency Electronic Apex Locators Assessed through Micro-Computed Tomographic Imaging.J Endod. 2023 May;49(5):487-495. doi: 10.1016/j.joen.2023.02.011. Epub 2023 Feb 26. J Endod. 2023. PMID: 36848949
-
In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi.J Endod. 2020 Feb;46(2):158-161. doi: 10.1016/j.joen.2019.10.035. Epub 2019 Dec 12. J Endod. 2020. PMID: 31839411
-
Comparative evaluation of the accuracy of six different apex locators in working length determination of molars using intraoral periapical radiographs: An in vivo study.J Conserv Dent Endod. 2024 Jul;27(7):695-700. doi: 10.4103/JCDE.JCDE_189_24. Epub 2024 Jul 8. J Conserv Dent Endod. 2024. PMID: 39262599 Free PMC article.
-
[Comparison of the precision of two electronic apical locators in teeth with simulated perforations].Rev Cient Odontol (Lima). 2021 Jun 21;9(2):e055.. doi: 10.21142/2523-2754-0902-2021-055. eCollection 2021 Apr-Jun. Rev Cient Odontol (Lima). 2021. PMID: 39687286 Free PMC article. Spanish.
-
Comparative Evaluation of Accuracy of Different Apex Locators: Propex IQ, Raypex 6, Root ZX, and Apex ID with CBCT and Periapical Radiograph-In Vitro Study.Int J Dent. 2021 May 4;2021:5563426. doi: 10.1155/2021/5563426. eCollection 2021. Int J Dent. 2021. PMID: 34054960 Free PMC article.
Cited by
-
Hybrid Approach to Manage Inflammatory Internal Root Resorption: A Case Report.Iran Endod J. 2025;20(1):e18. doi: 10.22037/iej.v20i1.46677. Epub 2025 Mar 8. Iran Endod J. 2025. PMID: 40206782 Free PMC article.
-
Influence of root canal taper on the measurement of two different electronic apex locators.J Clin Exp Dent. 2024 Jun 1;16(6):e733-e739. doi: 10.4317/jced.61352. eCollection 2024 Jun. J Clin Exp Dent. 2024. PMID: 39130363 Free PMC article.
-
Comparative Accuracy and Reliability of Three Electronic Apex Locators in Determining the Apical Constriction of Molar Canals: A Micro-CT Evaluation.J Clin Med. 2024 Sep 2;13(17):5199. doi: 10.3390/jcm13175199. J Clin Med. 2024. PMID: 39274411 Free PMC article.
References
-
- Rodríguez-Niklitschek C, Oporto V GH. Determinación de la Longitud de Trabajo en Endodoncia: Implicancias Clínicas de la Anatomía Radicular y del Sistema de Canales Radiculares. Int J Odontostomatol. 2014;8:177–83.
-
- Welk AR, Baumgartner JC, Marshall JG. An in vivo comparison of two frequency-based electronic apex locators. J Endod. 2003;29(8):497–500. - PubMed
-
- Katz A, Tamse A, Kaufman AY. Tooth length determination: a review. Oral Surg Oral Med Oral Pathol. 1991;72(2):238–42. - PubMed
-
- Tamse A, Kaffe I, Fishel D. Zygomatic arch interference with correct radiographic diagnosis in maxillary molar endodontics. Oral Surg Oral Med Oral Pathol. 1980;50(6):563–6. - PubMed
-
- Duran-Sindreu F, Gomes S, Stöber E, Mercadé M, Jané L, Roig M. In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants. Int Endod J. 2013;46(8):769–74. - PubMed
LinkOut - more resources