Optimal Positioning of Nasopharyngeal Temperature Probes in Infants and Children: A Prospective Cohort Study
- PMID: 36730063
- DOI: 10.1213/ANE.0000000000006312
Optimal Positioning of Nasopharyngeal Temperature Probes in Infants and Children: A Prospective Cohort Study
Abstract
Background: The nasopharynx is an easily accessible core-temperature monitoring site, but insufficient or excessive nasopharyngeal probe insertion can underestimate core temperature. Our goal was to estimate optimal nasopharyngeal probe insertion depth as a function of age.
Methods: We enrolled 157 pediatric patients who had noncardiac surgery with endotracheal intubation in 5 groups: (1) newborn to 6 months old, (2) infants 7 months to 1 year old, (3) children 13 to 23 months old, (4) children 2 to 5 years old, and (5) children 6 to 12 years old. A reference esophageal temperature probe was inserted at an appropriate depth based on each patient's height. A nasopharyngeal temperature probe was inserted from the naris at 10 cm in newborn and infants, 15 cm in children aged 1 to 5 years old, and 20 cm in children who were 6 years or older. The study nasopharyngeal probes were withdrawn 1, 2.5, or 2 cm (depending on age) 10 times at 5-minute intervals. Optimal probe insertion distances were defined by limits of agreement (LOAs) between nasopharyngeal and esophageal temperatures <0.5 °C.
Results: Optimal nasopharyngeal temperature probe position ranged from 6 to 10 cm in infants up to 6 months old, 7 to 8 cm in infants 7 to 12 months old, 7.5 to 12 cm in children 13 to 23 months old, and 10 to 12 cm in children aged 6 years and older. The 95% LOAs were <0.5 °C for all age categories except the 2- to 5-year-old group where the limits extended from -0.67 °C to 0.52 °C at 9 cm. At the optimal position within each age range, the bias (average nasopharyngeal-to-esophageal temperature difference) was ≤0.1 °C.
Conclusions: Nasopharyngeal thermometers accurately measure core temperature, but only when probes are inserted a proper distance, which varies with age. As with much in pediatrics, nasopharyngeal thermometer insertion depths should be age appropriate.
Copyright © 2022 International Anesthesia Research Society.
Conflict of interest statement
Conflicts of Interest: See Disclosures at the end of the article.
References
-
- Sessler DI. Temperature monitoring and perioperative thermoregulation. Anesthesiology. 2008;109:318–338.
-
- Sessler DI. Perioperative heat balance. Anesthesiology. 2000;92:578–596.
-
- Kurz A, Sessler DI, Lenhardt R. Perioperative normothermia to reduce the incidence of surgical-wound infection and shorten hospitalization. Study of wound infection and temperature group. N Engl J Med. 1996;334:1209–1215.
-
- Leslie K, Sessler DI, Bjorksten AR, Moayeri A. Mild hypothermia alters propofol pharmacokinetics and increases the duration of action of atracurium. Anesth Analg. 1995;80:1007–1014.
-
- Rajagopalan S, Mascha E, Na J, Sessler DI. The effects of mild perioperative hypothermia on blood loss and transfusion requirement. Anesthesiology. 2008;108:71–77.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials