Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Observational Study
. 2023 Apr;46(4):369-376.
doi: 10.1016/j.jfo.2022.09.034. Epub 2023 Feb 3.

[Comparison of available clinical and imaging tools to assess good positioning of a fluocinolone acetonide implant (Iluvien®) in the vitreous cavity after injection]

[Article in French]
Affiliations
Observational Study

[Comparison of available clinical and imaging tools to assess good positioning of a fluocinolone acetonide implant (Iluvien®) in the vitreous cavity after injection]

[Article in French]
E Jomaa et al. J Fr Ophtalmol. 2023 Apr.

Abstract

Introduction: Sustained-release corticosteroid implants are injected into the vitreous cavity using preloaded pens. The fluocinolone (FAc) implant is approximately half the size of the dexamethasone implant (Dex-I). It is simply introduced in the vitreous base rather than propelled into the vitreous cavity as is Dex-I. Verification of its positioning after injection is thus difficult by indirect ophthalmoscopy. The goal of our study is to compare the performance of available clinical and imaging tools to confirm the presence of the FAc in the vitreous cavity following injection.

Methods: Twelve eyes of 12 consecutive patients were included in a retrospective, single-center, observational study carried out at the Bordeaux University Hospital, France. All patients were injected with the FAc after pupil dilation, and presence of the implant was immediately checked by indirect biomicroscopy, wide-field retinography (Clarus®, Carl-Zeiss-Meditec, Dublin, CA, USA) and ultra-wide-field retinography (California®, Optos, Edinburgh, United-Kingdom). Seven days later, a B-mode ultrasonography (10MHz, AVISO, Quantel-medical, France) and an UBM ultrasonography (50MHz, AVISO, Quantel-medical, France) were performed.

Results: Indirect biomicroscopy and wide-field retinography detected 4/12 implants (33.3%). Ultra-wide-field retinophotography detected 6/12 implants (50%). All the implants seen using indirect biomicroscopy and wide-field retinography were also visualized with ultra-wide-field. B-mode ultrasonography showed 5/12 implants (41.6%) and UBM 9/12 implants (75%). Finally, one implant dislocated into the anterior chamber and was seen in the iridocorneal angle on gonioscopy.

Conclusion: Objective confirmation of the proper positioning of the FAc implant in the vitreous cavity is mandatory. If both indirect ophthalmoscopy and anterior examination fail to detect it, ultra-wide field retinography along with UBM ultrasonography, if necessary, appear to be the two best imaging modalities to use.