Automated synthesis and quality control of [68Ga]Ga-PentixaFor using the Gaia/Luna Elysia-Raytest module for CXCR4 PET imaging

Affiliations

Automated synthesis and quality control of [68Ga]Ga-PentixaFor using the Gaia/Luna Elysia-Raytest module for CXCR4 PET imaging

Thomas Daniel et al. EJNMMI Radiopharm Chem. .

Abstract

Background: [68Ga]Ga-PentixaFor is a promising radiotracer for positron emission tomography imaging of several human tumors overexpressing the chemokine receptor-4 (CXCR4). CXCR4 overexpression has been demonstrated in patients with hematologic malignancies, solid cancers, as well as cardiovascular pathologies of inflammatory origins. However, its radio synthesis is not yet fully developed in France, and existing methods do not use our type of synthesis module. Therefore, we aimed at developing a [68Ga]Ga-PentixaFor synthesis with Gaia/Luna Elysia-Raytest module to use it in clinical purpose.

Results: 12 syntheses were carried out by varying the temperature conditions and radiolabeling times, and led to choose specific labelling conditions with the Gaia/Luna Elysia-Raytest module: 97 °C, 4 min. The mean synthesis time of the 3 validation runs under good manufacturing practice (GMP) was 24 min 27 s (± 8 s), and the mean radiochemical yield was 87.0% [standard deviation (SD) 6.67%]. Different quality control parameters were also evaluated in accordance with European Pharmacopeia: radiochemical and radionuclidic purity, pH, sterility, stability and endotoxins levels. The average radiochemical purity was 99.1% (SD 0.25%) assessed by instant thin layer chromatography and 99.8% (SD 0.092%) assessed by high pressure liquid chromatography. average [68Ge] breakthrough was 1.48 × 10-5%, under the recommended level of 0.001%. We assessed the stability of the radiotracer up to 4 h at room temperature (no augmentation of the [68Ga] chloride in the final product, i.e. radiochemical purity (RCP) > 98.5%). The endotoxins levels were < 5 EU/mL, and the pH was 6.5 (same for the three syntheses).

Conclusion: The [68Ga]Ga-PentixaFor synthesis process developed on the Gaia/Luna Elysia-Raytest module has fulfilled all acceptance criteria for injectable radiopharmaceutical products regarding the European Pharmacopeia. The radiochemical purity, stability, efficacy, as well as the microbiological quality of the three GMP batches were found to be good. The robustness of the synthesis process may be suitable for multi-dose application in clinical settings.

Keywords: 68Ga; Automated synthesis; CXCR4; PET; PentixaFor; Radiopharmacy.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
Chemical Structure depiction of [68Ga]Ga-PentixaFor (Schottelius et al. 2017)
Fig. 2
Fig. 2
Flow-chart of the synthesis of the [68Ga]Ga-PentixaFor
Fig. 3
Fig. 3
Time course of RCP of [68Ga]Ga-Pentixafor during stability test assessed by iTLC
Fig. 4
Fig. 4
Time course of impurities (colloidal forms and free Ga3+ percentages) during stability test, assessed by iTLC
Fig. 5
Fig. 5
Time course of RCP of [68Ga]Ga-Pentixafor during stability test, assessed by HPLC

References

    1. Balkwill F. Cancer and the chemokine network. Nat Rev Cancer. 2004;4(7):540–550. doi: 10.1038/nrc1388. - DOI - PubMed
    1. Baz AHC, van de Wiel E, Groenewoud H, et al. CXCR4-directed [68Ga]Ga-PentixaFor PET/CT versus adrenal vein sampling performance: a study protocol for a randomised two-step controlled diagnoStic Trial Ultimately comparing hypertenSion outcome in primary aldosteronism (CASTUS) BMJ Open. 2022;12(8):e060779. doi: 10.1136/bmjopen-2022-060779. - DOI - PMC - PubMed
    1. Buck AK, Haug A, Dreher N, et al. Imaging of C-X-C motif chemokine receptor 4 expression in 690 patients with solid or hematologic neoplasms using 68Ga-pentixafor PET. J Nucl Med. 2022;63(11):1687–1692. doi: 10.2967/jnumed.121.263693. - DOI - PubMed
    1. Cui Y, Zhang Y, Ding J, et al. A rare aldosterone-producing adenoma detected by 68Ga-pentixafor PET-CT: a case report and literature review. Front Endocrinol. 2019;10:810. doi: 10.3389/fendo.2019.00810. - DOI - PMC - PubMed
    1. Demmer O, Gourni E, Schumacher U, Kessler H, Wester HJ. PET imaging of CXCR4 receptors in cancer by a new optimized ligand. ChemMedChem. 2011;6(10):1789–1791. doi: 10.1002/cmdc.201100320. - DOI - PMC - PubMed

LinkOut - more resources