Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Jan 27;24(3):2487.
doi: 10.3390/ijms24032487.

Comparison of Tandem Mass Spectrometry and the Fluorometric Method-Parallel Phenylalanine Measurement on a Large Fresh Sample Series and Implications for Newborn Screening for Phenylketonuria

Affiliations

Comparison of Tandem Mass Spectrometry and the Fluorometric Method-Parallel Phenylalanine Measurement on a Large Fresh Sample Series and Implications for Newborn Screening for Phenylketonuria

Dasa Perko et al. Int J Mol Sci. .

Abstract

Phenylketonuria (PKU) was the first disease to be identified by the newborn screening (NBS) program. Currently, there are various methods for determining phenylalanine (Phe) values, with tandem mass spectrometry (MS/MS) being the most widely used method worldwide. We aimed to compare the MS/MS method with the fluorometric method (FM) for measuring Phe in the dried blood spot (DBS) and the efficacy of both methods in the NBS program. The FM was performed using a neonatal phenylalanine kit and a VICTOR2TM D fluorometer. The MS/MS method was performed using a NeoBaseTM 2 kit and a Waters Xevo TQD mass spectrometer. The Phe values measured with the MS/MS method were compared to those determined by the FM. The cut-off value for the NBS program was set at 120 µmol/L for FM and 85 µmol/L for MS/MS. We analyzed 54,934 DBS. The measured Phe values varied from 12 to 664 µmol/L, with a median of 46 µmol/L for the MS/MS method and from 10 to 710 µmol/L, with a median of 70 µmol/L for the FM. The Bland-Altman analysis indicated a bias of -38.9% (-23.61 µmol/L) with an SD of 21.3% (13.89 µmol/L) when comparing the MS/MS method to the FM. The Phe value exceeded the cut-off in 187 samples measured with FM and 112 samples measured with MS/MS. The FM had 181 false positives, while the MS/MS method had 106 false positives. Our study showed that the MS/MS method gives lower results compared to the FM. Despite that, none of the true positives would be missed, and the number of false-positive results would be significantly lower compared to the FM.

Keywords: comparison; false positive; fluorometric method (FM); newborn screening (NBS); phenylalanine (Phe); phenylketonuria (PKU); recall rate; tandem mass spectrometry (MS/MS).

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
Scatter dot plot displaying the phenylalanine (Phe) values determined by the tandem mass spectrometry (MS/MS) and the fluorimetric method (FM). (A) The whole range of Phe values; (B) Phe values up to 150 µmol/L.
Figure 2
Figure 2
The Bland-Altman analysis of the tandem mass spectrometry (MS/MS) and the fluorimetric method (FM) for determining phenylalanine (Phe) concentrations in dried blood spots (DBS). The bias was −38.9%, with the lower (LLA) and upper (ULA) lines of agreement at −80.73% and 2.94%, respectively. (A) The whole range of Phe values; (B) Phe values up to 150 µmol/L.
Figure 3
Figure 3
Correlation between the tandem mass spectrometry (MS/MS) and the fluorimetric method (FM). The Spearman correlation coefficient (rs) was 0.49, with a 95% confidence interval of 0.483–0.496. (A) The whole range of phenylalanine (Phe) values; (B) Phe values up to 150 µmol/L.
Figure 4
Figure 4
Comparison of the tandem mass spectrometry (MS/MS) and the fluorimetric method (FM) for phenylketonuria (PKU) newborn screening (NBS); the number of analyzed dried blood spots (DBS) was 54,934.
Figure 5
Figure 5
Simulated number of recalls, false-positive and false-negative samples at different phenylalanine cut-off values.

References

    1. Pourfarzam M., Zadhoush F. Newborn screening for inherited metabolic disorders; news and views. J. Res. Med. Sci. 2013;18:801–808.
    1. Repič Lampret B., Remec Ž.I., Drole Torkar A., Žerjav Tanšek M., Šmon A., Koračin V., Čuk V., Perko D., Ulaga B., Jelovšek A.M., et al. Expanded newborn screening program in Slovenia using tandem mass spectrometry and confirmatory next generation sequencing genetic testing. Zdr. Varst. 2020;95:256–263. doi: 10.2478/sjph-2020-0032. - DOI - PMC - PubMed
    1. Kaufman S. The Phenylalanine Hydroxylating System. Adv. Enzymol. Relat. Areas Mol. Biol. 2006;67:77–264. doi: 10.1002/9780470123133.ch2. - DOI - PubMed
    1. Grošelj U., Tanšek M.Ž., Podkrajšek K.T., Battelino T. Genetske in klinične značilnosti bolnikov s fenilketonurijo v Sloveniji. Zdr. Vestn. 2013;82:767–777.
    1. Maxim L.D., Niebo R., Utell M.J. Screening tests: A review with examples. Inhal. Toxicol. 2014;26:811–828. doi: 10.3109/08958378.2014.955932. - DOI - PMC - PubMed