. 2023 Apr;34(4):833-840.
doi: 10.1111/jce.15861. Epub 2023 Feb 23.

Acute procedural efficacy and safety of a novel cryoballoon for the treatment of paroxysmal atrial fibrillation: Results from the POLAR ICE study

Affiliations
Multicenter Study

Acute procedural efficacy and safety of a novel cryoballoon for the treatment of paroxysmal atrial fibrillation: Results from the POLAR ICE study

Claire A Martin et al. J Cardiovasc Electrophysiol. 2023 Apr.

Abstract

Introduction: Pulmonary vein isolation (PVI) is well established as a primary treatment for atrial fibrillation (AF). The POLAR ICE study was designed to collect prospective real world data on the safety and effectiveness of the POLARxTM cryoballoon for PVI to treat paroxysmal AF.

Methods: POLAR ICE, a prospective, non-randomized, multicenter (international) registry (NCT04250714), enrolled 399 patients across 19 European centers. Procedural characteristics, such as time to isolation, cryoablations per pulmonary vein (PV), balloon nadir temperature, and occlusion grade were recorded. PVI was confirmed with entrance block testing.

Results: Data on 372 de novo PVI procedures (n = 2190 ablations) were collected. Complete PVI was achieved in 96.8% of PVs. Procedure and fluoroscopy times were 68.2 ± 24.6 and 15.6 ± 9.6 min, respectively. Left atrial dwell time was 46.6 ± 18.3 min. Grade 3 or 4 occlusion was achieved in 98.2% of PVs reported and 71.2% of PVs isolation required only a single cryoablation. Of 2190 cryoapplications, 83% had a duration of at least 120 s; nadir temperature of these ablations averaged -56.3 ± 6.5°C. There were 6 phrenic nerve palsy events, 2 of which resolved within 3 months of the procedure.

Conclusion: This real-world usage data on a novel cryoballoon suggests this device is effective, safe, and relatively fast in centers with cryoballoon experience. These data are comparable to prior POLARx reports and in keeping with reported data on other cryoballoons. Future studies should examine the long-term outcomes and the relationship between biophysical parameters and outcomes for this novel cryoballoon.

Keywords: cryoballoon ablation; paroxysmal atrial fibrillation; pulmonary vein isolation (PVI); safety; success rate; time to thaw.

PubMed Disclaimer

References

REFERENCES

    1. Hindricks G, Potpara T, Dagres N, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): the task force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021;42:373-498.
    1. Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275-e444.
    1. Kuck KH, Brugada J, Fürnkranz A, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med. 2016;374:2235-2245.
    1. Packer DL, Kowal RC, Wheelan KR, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation. JACC. 2013;61:1713-1723.
    1. Anic A, Lever N, Martin A, et al. Acute safety, efficacy, and advantages of a novel cryoballoon ablation system for pulmonary vein isolation in patients with paroxysmal atrial fibrillation: initial clinical experience. EP Europace. 2021;23:1237-1243.

Publication types

Associated data