Container closure integrity testing and process validation of closed system transfer devices for aseptic reconstitution of drug vials connected to fluid bags
- PMID: 36792349
- PMCID: PMC11265557
- DOI: 10.1136/ejhpharm-2022-003604
Container closure integrity testing and process validation of closed system transfer devices for aseptic reconstitution of drug vials connected to fluid bags
Abstract
Objectives: The closure integrity and process validation of closed system transfer devices (CSTDs) should be confirmed before implementation in clinical settings. We aimed to investigate the closure integrity and validate the aseptic procedure of two types of CSTDs by using a combination of the dye ingress test and a media fill test.
Methods: The dye ingress test with methylene blue was used for both CSTDs with 10 samples of drug vials of three brands. A media fill test was performed with both CSTDs (300 samples per CSTD, 150 carried out in a safety cabinet and 150 under non-classified environmental conditions).
Results: In all samples of both CSTDs, methylene blue was absent after visual inspection and spectrophotometric analysis. The nutrient media of one sample with CSTD A and none of the CSTD B samples were contaminated when reconstituted in a GMP grade A environment. Under non-classified environmental conditions, one sample of CSTD A and two samples of CSTD B were contaminated.
Conclusions: Both CSTDs connected to the drug vials met the terms of closure integrity by using the dye ingress test. The aseptic procedure of CSTD B was validated with the media fill test when reconstituted in a GMP grade A environment, but failed for CSTD A. Both CSTDs failed the media fill test when reconstituted under non-classified environmental conditions.
Keywords: chemical safety; drug incompatibility; pharmaceutical preparations; pharmacopoeia; safety.
© European Association of Hospital Pharmacists 2024. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: None declared.
Figures
Similar articles
-
Closed-system drug-transfer devices plus safe handling of hazardous drugs versus safe handling alone for reducing exposure to infusional hazardous drugs in healthcare staff.Cochrane Database Syst Rev. 2018 Mar 27;3(3):CD012860. doi: 10.1002/14651858.CD012860.pub2. Cochrane Database Syst Rev. 2018. PMID: 29582940 Free PMC article.
-
Recommendation of IV Dose Preparation Practices Using Closed System Transfer Devices (CSTD) for Accurate Dosing.AAPS J. 2025 May 16;27(4):93. doi: 10.1208/s12248-025-01084-z. AAPS J. 2025. PMID: 40380038
-
Validation of a customized combined chemical and microbiological test for qualification of aseptic preparation processes.J Oncol Pharm Pract. 2025 Jun 19:10781552251350633. doi: 10.1177/10781552251350633. Online ahead of print. J Oncol Pharm Pract. 2025. PMID: 40538112
-
Rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection.Cochrane Database Syst Rev. 2022 Jul 22;7(7):CD013705. doi: 10.1002/14651858.CD013705.pub3. Cochrane Database Syst Rev. 2022. PMID: 35866452 Free PMC article.
-
Advancing Parenteral Package Integrity Testing: Case Studies on IV Bags and Pre-Filled Syringes: Poster Presented at PDA Week 2025.PDA J Pharm Sci Technol. 2025 Aug 1;79(4):480-481. doi: 10.5731/pdajpst.2025.25429. PDA J Pharm Sci Technol. 2025. PMID: 40750368
References
-
- Pfizer . Summary of product characteristics meronem i.v. 500 mg, powder for solution for injection or infusion. 2019. Available: https://www.cbg-meb.nl [Accessed 21 Jan 2023].
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous