Review
. 2023 Mar 15:430:108709.
doi: 10.1016/j.heares.2023.108709. Epub 2023 Feb 1.

Effects of preloads on middle-ear transfer function and acoustic reflex in ossiculoplasty with PORP

Affiliations
Review

Effects of preloads on middle-ear transfer function and acoustic reflex in ossiculoplasty with PORP

Merlin Schär et al. Hear Res. .

Abstract

Introduction: Surgical outcomes in ossiculoplasty with partial ossicular replacement prostheses (PORPs) are greatly influenced by the amount of preload imposed on the PORP. In this study, the attenuation of the middle-ear transfer function (METF) was experimentally investigated for prosthesis-related preloads in different directions, with and without concurrent application of stapedial muscle tension. Different PORP designs were assessed to determine functional benefits of specific design features under preload conditions.

Methods: The experiments were performed on fresh-frozen human cadaveric temporal bones. The effect of preloads along different directions were experimentally assessed by simulating anatomical variance and postoperative position changes in a controlled setup. The assessments were performed for three different PORP designs featuring either a fixed shaft or ball joint and a Bell-type or Clip-interface. Further, the combined effect of the preloads towards the medial direction with tensional forces of the stapedial muscle was assessed. The METF was obtained via laser-Doppler vibrometry for each measurement condition.

Results: The preloads as well as the stapedial muscle tension primarily attenuated the METF between 0.5 and 4 kHz. The largest attenuations resulted from the preload towards the medial direction. The attenuation of the METF with stapedial muscle tension was reduced with concurrent PORP preloads. PORPs with a ball joint resulted in reduced attenuation only for preloads along the long axis of the stapes footplate. In contrast to the clip interface, the Bell-type interface was prone to lose coupling with the stapes head for preloads in the medial direction.

Conclusions: The experimental study of the preload effects indicates a direction-dependent attenuation of the METF, with the most pronounced effects resulting from preloads towards the medial direction. Based on the obtained results, the ball joint offers tolerance for angular positioning while the clip interface prevents PORP dislocations for preloads in lateral direction. At high preloads, the attenuation of the METF with stapedial muscle tension is reduced, which should be considered for the interpretation of postoperative acoustic reflex tests.

Keywords: Acoustic reflex; Annular ligament; Ossiculoplasty; Partial-ossicular replacement prosthesis; Preload; Stapedial muscle.