Rapid differential diagnosis of SARS-CoV-2, influenza A/B and respiratory syncytial viruses: Validation of a novel RT-PCR assay
- PMID: 36805601
- DOI: 10.1016/j.jcv.2023.105402
Rapid differential diagnosis of SARS-CoV-2, influenza A/B and respiratory syncytial viruses: Validation of a novel RT-PCR assay
Abstract
Background: Influenza and respiratory syncytial (RSV) viruses are expected to co-circulate with SARS-CoV-2 in the upcoming seasons and clinical differential diagnosis between them is difficult. Laboratory-based RT-PCR is a gold standard diagnostic method for influenza, RSV and SARS-CoV-2. The objective of this study was to estimate the diagnostic performance of a novel point-of-care RT-PCR assay STANDARD M10 Flu/RSV/SARS-CoV-2 (SD Biosensor) in a large number of clinical specimens with diversified (co)-infection patterns and viral loads.
Methods: This was a retrospective study, in which all samples were tested in both STANDARD M10 Flu/RSV/SARS-CoV-2 index and Allplex SARS-CoV-2/Respiratory Panel 1 (Seegene) reference kits. Samples with discordant results were further processed in a third resolver test (Resp-4-Plex, Abbott).
Results: A total of 1,019 naso-/oropharyngeal samples (50.3% positive for at least one virus) were processed in both STANDARD M10 Flu/RSV/SARS-CoV-2 and Allplex assays and the overall between-assay agreement was as high as 94.6%. Positive percent agreement of the STANDARD M10 Flu/RSV/SARS-CoV-2 was 100%, 96.6%, 97.3% and 99.4% for influenza A, B, RSV and SARS-CoV-2, respectively. The corresponding negative percent agreement was 99.7%. 100%, 100% and 98.4%, respectively. The expected positive and negative predictive values for all viruses were constantly above 96% in a reasonable range of disease prevalence.
Conclusions: STANDARD M10 Flu/RSV/SARS-CoV-2 is a reliable RT-PCR assay able to detect influenza A, influenza B, RSV and SARS-CoV-2 in one hour or less, fostering a rapid differential diagnosis of common respiratory viruses.
Keywords: Diagnostic accuracy; Influenza; RSV; SARS-CoV-2; point-of-care testing; rapid RT-PCR.
Copyright © 2023. Published by Elsevier B.V.
Conflict of interest statement
Declaration of Competing Interest A.D. was previously a full-time employee of Seqirus, a pharmaceutical company that manufactures and markets influenza vaccines, and received speaker fees from SD Biosensor. The other authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous