Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Jan 30;13(2):384.
doi: 10.3390/life13020384.

Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor-A Clinical Validation Study

Affiliations

Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor-A Clinical Validation Study

Liat Iacubovici et al. Life (Basel). .

Abstract

Introduction: Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.

Methods: Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each participant. Measurements were compared using Bland-Altman analysis.

Results: In total, 29 participants completed the study and provided 144 distinct pairs of pressure measurements that were analyzed. A positive correlation between the two methods was found (R2 = 0.93). There was good agreement between the methods, with a mean bias (95% CI) of -0.4 (-0.6, -0.1) mmHg and a standard deviation of 1.3 mmHg, which was statistically significant but of no clinical importance. The limits of agreement (where 95% of the differences are expected to fall) were -2.9 and 2.2 mmHg. The proportional error was statistically insignificant (p = 0.85), suggesting a constant agreement between the methods across the range of values tested. The percentage error was 10.7%.

Conclusions: Continuous IAP measurements using the novel monitor performed well in the clinical setup of controlled intra-abdominal hypertension across the evaluated range of pressures. Further studies should expand the range to more pathological values.

Keywords: abdominal compartment syndrome; foley manometer; intra-abdominal hypertension; intra-abdominal pressure monitoring.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Figures

Figure 1
Figure 1
Study system design (not to scale). A—Insufflator—allows setting a pre-defined abdominal pressure during the study. B—Study device control unit—mounted on a pole or bed-rail. Can be placed anywhere near the bed or move with the patient. C—Tube from CO2 insufflator to the surgical port. D—Classic fluid column manometer measures intra-vesical pressure via a Foley catheter. E—Surgical port into the peritoneal cavity. F—Mid-axillary line (reference level). G—Urinary bladder. H—Urine collection bag. I—Study device disposable measurement unit, located between the Foley catheter and urine bag, and transmitting pressures to the controller.
Figure 2
Figure 2
Study flow-chart.
Figure 3
Figure 3
Scatter-plot of acceptance between the two measurement methods.
Figure 4
Figure 4
Bland–Altman analysis. The Y-axis shows the difference between measurements made by the study device minus those of the foley manometer, and the X-axis presents the means of measurements via both methods. The red line represents the mean bias (-0.4 mmHg), and green lines represent the upper and lower limits of agreement (2.2 and -2.9 mmHg, respectively).

Similar articles

Cited by

References

    1. Al-Abassi A.A., al Saadi A.S., Ahmed F. Is Intra-Bladder Pressure Measurement a Reliable Indicator for Raised Intra-Abdominal Pressure? A Prospective Comparative Study. BMC Anesthesiol. 2018;18:69. doi: 10.1186/s12871-018-0539-z. - DOI - PMC - PubMed
    1. Milanesi R., Caregnato R.C.A. Intra-Abdominal Pressure: An Integrative Review. Einstein. 2016;14:423–430. doi: 10.1590/S1679-45082016RW3088. - DOI - PMC - PubMed
    1. Parsak C.K., Seydaoglu G., Sakman G., Acarturk T.O., Karakoc E., Hanta I., Alparslan A.H., Satar S. Abdominal Compartment Syndrome: Current Problems and New Strategies. World J. Surg. 2008;32:13–19. doi: 10.1007/s00268-007-9286-x. - DOI - PubMed
    1. De Laet I.E., Ravyts M., Vidts W., Valk J., de Waele J.J., Malbrain M.L.N.G. Current Insights in Intra-Abdominal Hypertension and Abdominal Compartment Syndrome: Open the Abdomen and Keep It Open! Langenbecks Arch. Surg. 2008;393:833–847. doi: 10.1007/s00423-008-0347-x. - DOI - PubMed
    1. Bodnar Z. Polycompartment Syndrome—Intra-Abdominal Pressure Measurement. Anaesthesiol. Intensive Ther. 2019;51:316–322. doi: 10.5114/ait.2019.87474. - DOI - PubMed