Comparative Study
. 2023 Apr:161:110734.
doi: 10.1016/j.ejrad.2023.110734. Epub 2023 Feb 9.

Liver fat quantification in photon counting CT in head to head comparison with clinical MRI - First experience

Affiliations
Comparative Study

Liver fat quantification in photon counting CT in head to head comparison with clinical MRI - First experience

Fides Regina Schwartz et al. Eur J Radiol. 2023 Apr.

Abstract

Purpose: To compare liver fat quantification between MRI and photon-counting CT (PCCT).

Method: A cylindrical phantom with inserts containing six concentrations of oil (0, 10, 20, 30, 50 and 100%) and oil-iodine mixtures (0, 10, 20, 30 and 50% fat +3 mg/mL iodine) was imaged with a PCCT (NAEOTOM Alpha) and a 1.5 T MRI system (MR 450w, IDEAL-IQ sequence), using clinical parameters. An IRB-approved prospective clinical evaluation included 12 obese adult patients with known fatty liver disease (seven women, mean age: 61.5 ± 13 years, mean BMI: 30.3 ± 4.7 kg/m2). Patients underwent a same-day clinical MRI and PCCT of the abdomen. Liver fat fractions were calculated for four segments (I, II, IVa and VII) using in- and opposed-phase on MRI ((Meanin - Meanopp)/2*Meanin) and iodine-fat, tissue decomposition analysis in PCCT (Syngo.Via VB60A). CT and MRI Fat fractions were compared using two-sample t-tests with equal variance. Statistical analysis was performed using RStudio (Version1.4.1717).

Results: Phantom results showed no significant differences between the known fat fractions (P = 0.32) or iodine (P = 0.6) in comparison to PCCT-measured concentrations, and no statistically significant difference between known and MRI-measured fat fractions (P = 0.363). In patients, the mean fat signal fraction measured on MRI and PCCT was 13.1 ± 9.9% and 12.0 ± 9.0%, respectively, with an average difference of 1.1 ± 1.9% between the modalities (P = 0.138).

Conclusion: First experience shows promising accuracy of liver fat fraction quantification for PCCT in obese patients. This method may improve opportunistic screening for CT in the future.

Keywords: Computed tomography; Hepatic steatosis; Magnetic resonance imaging; Photon counting; Quantification.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.