Guided growth: angular deformity correction through temporary hemiepiphysiodesis with a novel flexible staple (FlexTack)
- PMID: 36854319
- DOI: 10.1302/0301-620X.105B3.BJJ-2022-0857.R1
Guided growth: angular deformity correction through temporary hemiepiphysiodesis with a novel flexible staple (FlexTack)
Abstract
Temporary hemiepiphysiodesis (HED) is applied to children and adolescents to correct angular deformities (ADs) in long bones through guided growth. Traditional Blount staples or two-hole plates are mainly used for this indication. Despite precise surgical techniques and attentive postoperative follow-up, implant-associated complications are frequently described. To address these pitfalls, a flexible staple was developed to combine the advantages of the established implants. This study provides the first results of guided growth using the new implant and compares these with the established two-hole plates and Blount staples. Between January 2013 and December 2016, 138 patients (22 children, 116 adolescents) with genu valgum or genu varum were treated with 285 flexible staples. The minimum follow-up was 24 months. These results were compared with 98 patients treated with 205 two-hole plates and 92 patients treated with 535 Blount staples. In long-standing anteroposterior radiographs, mechanical axis deviations (MADs) were measured before and during treatment to analyze treatment efficiency. The evaluation of the new flexible staple was performed according to the idea, development, evaluation, assessment, long-term (IDEAL) study framework (Stage 2a). Overall, 79% (109/138) of patients treated with flexible staples achieved sufficient deformity correction. The median treatment duration was 16 months (interquartile range (IQR) 8 to 21). The flexible staples achieved a median MAD correction of 1.2 mm/month/HED site (IQR 0.6 to 2.0) in valgus deformities and 0.6 mm/month/HED site (IQR 0.2 to 1.5) in varus deformities. Wound infections occurred in 1%, haematomas and joint effusions in 4%, and implant-associated complications in 1% of patients treated with flexible staples. Valgus AD were corrected faster using flexible staples than two-hole plates and Blount staples. Furthermore, the median MAD after treatment was lower in varus and valgus AD, fewer implant-associated complications were detected, and reduced implantation times were recorded using flexible staples. The flexible staple seems to be a viable option for guided growth, showing comparable or possibly better results regarding correction speed and reducing implant-associated complications. Further comparative studies are required to substantiate these findings.
© 2023 The British Editorial Society of Bone & Joint Surgery.
Conflict of interest statement
R. Roedl is a paid consultant (less than $10,000) of Merete (Berlin, Germany) and has received gratifications for the development of PediatrOS FlexTack/RigidTack during the study period outside the submitted work. The other authors have no conflicts of interest.
References
-
- Ghanem I , Karam JA , Widmann RF . Surgical epiphysiodesis indications and techniques: update . Curr Opin Pediatr . 2011 ; 23 ( 1 ): 53 – 59 . 10.1097/MOP.0b013e32834231b3 21169837
-
- Vogt B , Schiedel F , Rödl R . Guided growth in children and adolescents. Correction of leg length discrepancies and leg axis deformities . Orthopade . 2014 ; 43 ( 3 ): 267 – 284 .
-
- Stevens PM . Guided growth: 1933 to the present . Strateg Trauma Limb Reconstr . 2006 ; 1 ( 1 ): 29 – 35 .
-
- Stevens PM . Guided growth for angular correction: a preliminary series using a tension band plate . J Pediatr Orthop . 2007 ; 27 ( 3 ): 253 – 259 . 10.1097/BPO.0b013e31803433a1 17414005
-
- Wiemann JM , Tryon C , Szalay EA . Physeal stapling versus 8-plate hemiepiphysiodesis for guided correction of angular deformity about the knee . J Pediatr Orthop . 2009 ; 29 ( 5 ): 481 – 485 . 10.1097/BPO.0b013e3181aa24a8 19568021