. 1987 Dec;33(12):2262-7.

Bias and precision of cholesterol analysis by physician's office analyzers

Affiliations
  • PMID: 3690842

Bias and precision of cholesterol analysis by physician's office analyzers

T R Koch et al. Clin Chem. 1987 Dec.

Abstract

We studied the bias and precision of serum cholesterol analysis by physician's office analyzers. Total imprecision (CV range, %) for analysis of serum pools was: Abbott Vision 1.5%-1.9%; Ames Seralyzer 3.9%-4.5%; BMD Reflotron 2.3%-3.8%; Chrometrics Cholesterol Test System 2.3%-2.8%; Kodak DT-60 1.6%-2.2%. The Ames Seralyzer exhibited an excessive between-run component of variation. We collected, from 109 volunteers, samples of venous serum, heparin-treated whole blood, heparin-treated plasma, and fingerstick whole blood, and analyzed each type (where possible) with each system; serum was also analyzed in duplicate (by a proposed Reference Method) at the Centers for Disease Control. For assays with serum, the BMD Reflotron and Kodak DT-60 exhibited negative bias. All systems gave lower results for plasma and whole blood than for serum. All systems except the Kodak DT-60 were less precise for analysis of patients' sera than for analysis of serum pools; between-specimen variables may influence the results of these systems.

PubMed Disclaimer

Publication types

LinkOut - more resources