Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Mar 1;12(3):15.
doi: 10.1167/tvst.12.3.15.

Noncontact Acoustic Micro-Tapping Optical Coherence Elastography for Quantification of Corneal Anisotropic Elasticity: In Vivo Rabbit Study

Affiliations

Noncontact Acoustic Micro-Tapping Optical Coherence Elastography for Quantification of Corneal Anisotropic Elasticity: In Vivo Rabbit Study

Mitchell A Kirby et al. Transl Vis Sci Technol. .

Abstract

Purpose: The purpose of this study was to demonstrate accurate measurement of corneal elastic moduli in vivo with noncontact and noninvasive optical coherence elastography.

Methods: Elastic properties (in-plane Young's modulus, E, and both in-plane, μ, and out-of-plane, G, shear moduli) of rabbit cornea were quantified in vivo using noncontact dynamic acoustic micro-tapping optical coherence elastography (AµT-OCE). The intraocular pressure (IOP)-dependence of measured mechanical properties was explored in extracted whole globes following in vivo measurement. A nearly incompressible transverse isotropic (NITI) model was used to reconstruct moduli from AµT-OCE data. Independently, cornea elastic moduli were also measured ex vivo with traditional, destructive mechanical tests (tensile extensometry and shear rheometry).

Results: Our study demonstrates strong anisotropy of corneal elasticity in rabbits. The in-plane Young's modulus, computed as E = 3μ, was in the range of 20 MPa to 44 MPa, whereas the out-of-plane shear modulus was in the range of 34 kPa to 261 kPa. Both pressure-dependent ex vivo OCE and destructive mechanical tests performed on the same samples within an hour of euthanasia strongly support the results of AµT-OCE measurements.

Conclusions: Noncontact AµT-OCE can noninvasively quantify cornea anisotropic elastic properties in vivo.

Translational relevance: As optical coherence tomography (OCT) is broadly accepted in ophthalmology, these results suggest the potential for rapid translation of AµT-OCE into clinical practice. In addition, AµT-OCE can likely improve diagnostic criteria of ectatic corneal diseases, leading to early diagnosis, reduced complications, customized surgical treatment, and personalized biomechanical models of the eye.

PubMed Disclaimer

Conflict of interest statement

Disclosure: M.A. Kirby, None; G. Regnault, None; I. Pelivanov, None; M. O'Donnell, None; R.K. Wang, None; T.T. Shen, None

Figures

Figure 1.
Figure 1.
(a) Schematic of AµT-OCE system (b) in vivo imaging of rabbit cornea (c) ex vivo whole globe with needle insertion to control IOP.
Figure 2.
Figure 2.
(a) Measured corneal surface vibrations in a whole globe rabbit sample inflated to a pressure of 7 mmHg. (b) Filtered and windowed x-t plot. (c) Best fit of A0 mode plotted on top of k-f plot. (d) Optimization function showing best-fit and uncertainty around G and (e) µ. Uncertainty intervals (red lines) calculated for the representative example where standard deviation in gNITI was 0.3%. The best-fit solution provided G = 20.5 kPa, with uncertainty of 19.8 kPa – 21.3 kPa, and µ =  4.0 MPa, with uncertainty of 2.2 MPa – 8.7 MPa.
Figure 3.
Figure 3.
Tono-Pen measured IOP value (mmHg) as the internal pressure was raised using a lifted water bath. Colored dots and lines correspond with each sample, black squares and error bars denote mean and standard deviation, respectively, and the black solid line corresponds with the best-fit linear function (Equation 7). The dotted black line is the one-to-one (slope = 1) line for visualization.
Figure 4.
Figure 4.
Summary of elasticity measurements performed in vivo and ex vivo. The squares denote in vivo values, and the triangles are the mean of ex vivo measurements. Each black dot is the OCE measured ex vivo modulus for a single sample at the associated pressure (all results shown in Supplementary Materials).
Figure 5.
Figure 5.
The mean and standard deviation for all corneas measured via OCE in vivo and ex vivo, as well as by destructive mechanical testing. (a) mean shear, G, modulus and (b) Young's, E, modulus for: in vivo OCE at the fixed physiological IOP (measured with the Tono-Pen and equal to 11 mmHg on average after the correction presented in Fig. 3), ex vivo OCE for three different ranges of IOP indicated in the legend, the mean rheometry value of the storage modulus at 16 Hz, and tensile testing value of the tangential modulus at 10% strain. Standard deviation corresponds to the deviation of measured moduli across the population of nine cornea samples.

Update of

Similar articles

Cited by

References

    1. Petsche SJ, Chernyak D, Martiz J, Levenston ME, Pinsky PM. Depth-dependent transverse shear properties of the human corneal stroma. Investig Ophthalmol Vis Sci. 2012; 53: 873–880. - PMC - PubMed
    1. Hatami-Marbini H. Viscoelastic shear properties of the corneal stroma. J Biomech. 2014; 47: 723–728. - PubMed
    1. Sloan SR, Khalifa YM, Buckley MR. The location- and depth-dependent mechanical response of the human cornea under shear loading. Investig Ophthalmol Vis Sci. 2014; 55: 7919–7924. - PubMed
    1. Zeng Y, Yang J, Huang K, Lee Z, Lee X. A comparison of biomechanical properties between human and porcine cornea. J Biomech. 2001; 34: 533–537. - PubMed
    1. Wollensak G, Spoerl E, Seiler T. Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking. J Cataract Refract Surg. 2003; 29: 1780–1785. - PubMed

Publication types