. 2023 Jun;112(6):1586-1594.
doi: 10.1016/j.xphs.2023.03.009. Epub 2023 Mar 17.

Effect of Various Silicone Oil And Tungsten Levels on the Stability of a Monoclonal Antibody in Nine Commercially Available Prefilled Syringes

Affiliations

Effect of Various Silicone Oil And Tungsten Levels on the Stability of a Monoclonal Antibody in Nine Commercially Available Prefilled Syringes

Markela Ibo Murphy et al. J Pharm Sci. 2023 Jun.

Abstract

Prefilled syringes are widely used as a primary container for therapeutic proteins because they are more convenient than glass vials. The stability of biologic molecules can be affected by different syringe materials and techniques, such as silicone oil levels and coating method, amount of tungsten remaining in the glass barrel after using a tungsten pin to create the needle hole, and end of the syringe, which can be Luer locked or pre-staked with a needle. We investigated the impact of these parameters by using a monoclonal antibody to collect the antibody's stability profile and the prefilled syringes' functionality data. Silicone oil levels had no impact on aggregation levels, and particle counts were lowest for silicone oil-free syringes. Functionality performance was similar and did not change throughout all stability time points for all syringe configurations. The break-loose force for Ompi syringes was initially lower and increased over time to align with those of the other configurations, all of which remained well below 25 N. Tungsten contaminants and agitation stress from shipping studies did not impact quality attributes. This work can help guide the development of similar products in prefilled syringes to ensure selection of the primary container that provides adequate stability for the protein, as well as maintain the desired functionality features over the shelf life of the drug product.

Keywords: Glide force; Particle size; Prefilled syringe; Protein formulation; Protein stability; Silicone oil; Tungsten.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

LinkOut - more resources