Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Mar 20;25(1):19.
doi: 10.1186/s12968-023-00926-z.

Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

Affiliations

Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

Constantin-Cristian Topriceanu et al. J Cardiovasc Magn Reson. .

Abstract

Introduction: A long T2 relaxation time can reflect oedema, and myocardial inflammation when combined with increased plasma troponin levels. Cardiovascular magnetic resonance (CMR) T2 mapping therefore has potential to provide a key diagnostic and prognostic biomarkers. However, T2 varies by scanner, software, and sequence, highlighting the need for standardization and for a quality assurance system for T2 mapping in CMR.

Aim: To fabricate and assess a phantom dedicated to the quality assurance of T2 mapping in CMR.

Method: A T2 mapping phantom was manufactured to contain 9 T1 and T2 (T1|T2) tubes to mimic clinically relevant native and post-contrast T2 in myocardium across the health to inflammation spectrum (i.e., 43-74 ms) and across both field strengths (1.5 and 3 T). We evaluated the phantom's structural integrity, B0 and B1 uniformity using field maps, and temperature dependence. Baseline reference T1|T2 were measured using inversion recovery gradient echo and single-echo spin echo (SE) sequences respectively, both with long repetition times (10 s). Long-term reproducibility of T1|T2 was determined by repeated T1|T2 mapping of the phantom at baseline and at 12 months.

Results: The phantom embodies 9 internal agarose-containing T1|T2 tubes doped with nickel di-chloride (NiCl2) as the paramagnetic relaxation modifier to cover the clinically relevant spectrum of myocardial T2. The tubes are surrounded by an agarose-gel matrix which is doped with NiCl2 and packed with high-density polyethylene (HDPE) beads. All tubes at both field strengths, showed measurement errors up to ≤ 7.2 ms [< 14.7%] for estimated T2 by balanced steady-state free precession T2 mapping compared to reference SE T2 with the exception of the post-contrast tube of ultra-low T1 where the deviance was up to 16 ms [40.0%]. At 12 months, the phantom remained free of air bubbles, susceptibility, and off-resonance artifacts. The inclusion of HDPE beads effectively flattened the B0 and B1 magnetic fields in the imaged slice. Independent temperature dependency experiments over the 13-38 °C range confirmed the greater stability of shorter vs longer T1|T2 tubes. Excellent long-term (12-month) reproducibility of measured T1|T2 was demonstrated across both field strengths (all coefficients of variation < 1.38%).

Conclusion: The T2 mapping phantom demonstrates excellent structural integrity, B0 and B1 uniformity, and reproducibility of its internal tube T1|T2 out to 1 year. This device may now be mass-produced to support the quality assurance of T2 mapping in CMR.

Keywords: Phantom; Quality control; T1 mapping; T2 mapping.

PubMed Disclaimer

Conflict of interest statement

The views expressed in this article are those of the authors who declare that they have no conflict of interest (financial or non-financial). W.P. is an employee of Resonance Health which is a medium commercial corporation.

Figures

Fig. 1
Fig. 1
i Schematic (not to scale) showing the internal and external phantom structure. ii Phantom front view showing isocenter line and liquid crystal display thermometer. HDPE = high-density polyethylene; PVC = polyvinyl chloride; PC = polycarbonate
Fig. 2
Fig. 2
i T1 and T2 (in ms) in the T2 phantom (n = 1) as measured at 1.5 T and 3 T: slow scan reference T1 obtained using inversion recovery (IR) gradient echo (GRE) (purple) and reference T2 using single echo (SE) (orange); T1 via modified Look-Locker inversion recovery (MOLLI) T1 mapping (green) and T2 via balanced steady state free precession (bSSFP) T2 mapping (blue); T2 obtained by the manufacturer in Australia using a 1.4 T Bruker minispec relaxometer at 22 °C (red). Tube arrangement is such that the more temperature-dependent and therefore unstable long-T1 tubes are away from the corners and towards the middle of the 3 × 3 array. ii Exemplar T2 and T1 maps on a Siemens 3 T Prisma clinical CMR scanner. ID = tube identity
Fig. 3
Fig. 3
i B0 field homogeneity across the nine phantom compartments as a measure of off-resonance in Hertz (Hz) at 1.5 T (blue) and 3 T (green) are shown (bottom). The associated B0 field maps with the field of view capturing the whole phantom at 1.5 T and 3 T are also presented (top–tube positions are overlaid in red). ii) B1 field homogeneity across the nine phantom compartments as a measure of the FA (in degrees) at 1.5 T (red) and 3 T (blue) are shown (bottom). These represent small shifts in FA or frequency (e.g.,10 Hz = 0.08 ppm at 3 T) and should not be regarded as significantly different between the tube compartments. As expected, the variation of relative FA is larger at 3 T (0.590–0.656) compared to 1.5 T (0.849–0.866). The associated B1 field maps of at 1.5 T and 3 T are also presented (top–tube positions are overlaid in red). FA flip angle. Other abbreviations as in Fig. 2
Fig. 4
Fig. 4
Comparison of T2 obtained by reference (long-TR) SE sequences (yellow) versus bSSFP T2 mapping (grey) at 1.5 T (Siemens Aera, left) and 3 T (Siemens Prisma, right) on the final phantom (n = 1) at baseline. TR repetition time. Other abbreviations as in Fig. 2
Fig. 5
Fig. 5
Temperature tests carried out at PTB–German Physikalisch-Technische Bundesanstalt (left)–using a 3 T Siemens Magnetom Verio (VB17) and a 12-channel head coil and at NIST–US National Institute of Standards and Technology (right)–using an Agilent 3 T small bore scanner. T1 was measured by IRSE, and T2 by SE. The measurements were performed on the final phantom (n = 1) at baseline. TE echo time. Other abbreviations as in Fig. 2
Fig. 6
Fig. 6
Short-term reproducibility of T2 at 1.5 T (left) and 3 T (right) acquired using T2 mapping bSSFP repeated 3 times in each of the final prototypes #Ci (at a temperature of 22°) (n = 1) and #Cii (at 21°) (n = 1) manufactured months apart, from independent stock solutions. All these scans were performed on the same day with independent placement of phantom and shims. Coefficients of variation (CoV) of T2 are shown per tube and were all < 1% in the absence of temperature correction. CoV for T1 using 3 MOLLI repeats are not shown here but were also < 1% for both prototypes (1.5 T range: 0.13–0.94%; 3 T range: 0.03–0.38%)

References

    1. Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S, et al. Myocardial T1 mapping and extracellular volume quantification: A Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92–92. doi: 10.1186/1532-429X-15-92. - DOI - PMC - PubMed
    1. Topriceanu C-C, Pierce I, Moon J, Captur G. T 2 and T 2⁎ mapping and weighted imaging in cardiac MRI. In: Magnetic Resonance Imaging; 2022:15–32. - PubMed
    1. Roy C, Slimani A, De Meester C, Amzulescu M, Pasquet A, Vancraeynest D, Vanoverschelde J-L, Pouleur A-C, Gerber BL. Age and sex corrected normal reference values of T1, T2 T2∗and ECV in healthy subjects at 3T CMR. J Cardiovasc Magn Reson. 2017;19:72–72. doi: 10.1186/s12968-017-0371-5. - DOI - PMC - PubMed
    1. Granitz M, Motloch LJ, Granitz C, Meissnitzer M, Hitzl W, Hergan K, Schlattau A. Comparison of native myocardial T1 and T2 mapping at 1.5T and 3T in healthy volunteers: Reference values and clinical implications. Wien Klin Wochenschr. 2018;131:143–155. doi: 10.1007/s00508-018-1411-3. - DOI - PMC - PubMed
    1. Wiesmueller M, Wuest W, Heiss R, Treutlein C, Uder M, May MS. Cardiac T2 mapping: Robustness and homogeneity of standardized in-line analysis. J Cardiovasc Magn Reson. 2020;22:39–39. doi: 10.1186/s12968-020-00619-x. - DOI - PMC - PubMed

Publication types