Morbidity and mortality of central venous catheter hemodialysis: a review of 10 years' experience
- PMID: 3696329
- DOI: 10.1159/000184523
Morbidity and mortality of central venous catheter hemodialysis: a review of 10 years' experience
Abstract
The morbidity and mortality of hemodialysis by internal central venous catheterization in the subclavian and internal jugular positions are reviewed. A follow-up study was performed in our unit over 10 years (786 catheterizations). The most frequent complications were inadequate flow (7.6%) inadvertent withdrawal (5.6%) and bacteremia (5.1%). The overall complication rate was 27.2%. Kinking (p less than 0.001), bleeding (p less than 0.01) and bacteremia (p less than 0.05) occurred more frequently in patients with chronic renal failure, compared to patients with acute renal failure. Inadvertent withdrawal was the only complication observed more frequently in the internal jugular than in the subclavian position (10.8 vs. 4.3%; p less than 0.01). Bacteremia occurred more frequently after prolonged periods of catheterization (greater than 10 days). No fatal complications were observed. To obtain a more accurate idea about mortality, two supplementary large groups were studied: a review of 11 published series (1,542 catheterizations) and a questionnaire-based survey of 16 dialysis centers (approximately 4,000 catheterizations). Six fatalities were registered: 1 due to septicemia (in the literature review) and 3 due to traumatic perforation of the cardiac or the vessel wall, 1 to septicemia and 1 to air embolism (in the questionnaire-based survey). Based on the three different groups studied, the mortality of catheter dialysis could be estimated to be between 0 and 1.25/1,000 catheterizations.
Similar articles
-
Risk factor analysis for long-term tunneled dialysis catheter-related bacteremias.Nephron. 2002 Jul;91(3):399-405. doi: 10.1159/000064279. Nephron. 2002. PMID: 12119469
-
[Recirculation of urea and dialysis efficiency using dual-lumen dialysis catheters in various locations: may the venous lumen of the catheter be used as the arterial lumen and vice versa?].Lijec Vjesn. 2003 Jan-Feb;125(1-2):1-5. Lijec Vjesn. 2003. PMID: 12812016 Croatian.
-
Tunneled internal jugular catheters in adult patients: comparison of outcomes in hemodialysis versus infusion catheters.Acta Radiol. 2007 Jul;48(6):613-9. doi: 10.1080/02841850701342104. Acta Radiol. 2007. PMID: 17611867
-
Carotid-jugular arteriovenous fistula: a complication of temporary hemodialysis catheter.Am J Nephrol. 1995;15(4):332-6. doi: 10.1159/000168859. Am J Nephrol. 1995. PMID: 7573193 Review.
-
Transverse cervical artery pseudoaneurysm: a rare complication of internal jugular vein cannulation.Am J Nephrol. 2000 Nov-Dec;20(6):476-82. doi: 10.1159/000046203. Am J Nephrol. 2000. PMID: 11146316 Review.
Cited by
-
Incidence and challenges in management of hemodialysis catheter-related infections.Sci Rep. 2022 Nov 29;12(1):20536. doi: 10.1038/s41598-022-23787-5. Sci Rep. 2022. PMID: 36446808 Free PMC article.
-
Percutaneous catheterization of the internal jugular vein for hemodialysis.Korean J Intern Med. 2001 Dec;16(4):242-6. doi: 10.3904/kjim.2001.16.4.242. Korean J Intern Med. 2001. PMID: 11855153 Free PMC article.
-
Risk factors associated with hemodialysis central venous catheter malfunction; a retrospective analysis of a randomized controlled trial.Can J Kidney Health Dis. 2014 Jul 8;1:15. doi: 10.1186/2054-3581-1-15. eCollection 2014. Can J Kidney Health Dis. 2014. PMID: 25960883 Free PMC article.
-
Non-imaging assisted insertion of un-cuffed, non-tunneled internal jugular venous catheters for hemodialysis: Safety and utility in modern day world.Biomed J. 2016 Aug;39(4):283-288. doi: 10.1016/j.bj.2015.12.004. Epub 2016 Sep 28. Biomed J. 2016. PMID: 27793271 Free PMC article.
-
[Sticking guide wire. Problems with a high-flow catheter].Anaesthesist. 2011 Jul;60(7):641-6. doi: 10.1007/s00101-011-1870-7. Epub 2011 Mar 16. Anaesthesist. 2011. PMID: 21424310 German.
MeSH terms
LinkOut - more resources