High-Throughput and Automated Detection of HLA-B*27 Using the LabTurboTM AIO System

Affiliations

High-Throughput and Automated Detection of HLA-B*27 Using the LabTurboTM AIO System

Yung-Che Chou et al. Biomedicines. .

Abstract

The adoption of an automated system can decrease the hands-on time requirements in a clinical laboratory setting. For the detection of HLA-B*27, implementing a high-throughput and fully automated system has several advantages over using manual methods. Therefore, this study aimed to evaluate automation efficiency for the detection of HLA-B*27. Peripheral blood samples were obtained from 50 participants, and DNA was isolated from these samples. A Pharmigene PG27 detection kit was used for the qualitative detection of HLA-B*27. The performances of the semi-automated and fully automated LabTurboTM AIO systems in the detection of HLA-B*27 were compared. The mean absorbance (optical density) values for the MaelstromTM 8 and LabTurboTM AIO systems were found to be 1.88 and 1.9, respectively. The housekeeping gene was amplified and quantified using a real-time PCR assay across all DNA extracts to check the quality of the extracted human DNA. The results were expressed as the cycle threshold (Ct) values for all DNA extracts from both platforms. The mean Ct values for the Roche Cobas z480 and LabTurboTM AIO systems were found to be 22.7 and 20.4, respectively. This study demonstrated that the semi-automated method and the LabTurboTM AIO system yield consistent results for the detection of HLA-B*27. However, compared to the semi-automated method, the LabTurboTM AIO system provides standardized procedures, avoids manual handling, and improves turnaround time.

Keywords: fully automated; human leukocyte antigen B27; semi-automated.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
Ct value of the distribution of target genes of MaelstromTM 8 and LabTurboTM AIO system when analyzing the clinical samples. Box plots illustrate the median value and interquartile range, and whiskers represent the nearest value to 1.5 times IQR. Outliers are plotted separately. IC (internal control), PG27 (HLA-B*27).

References

    1. Yonekawa T., Watanabe H., Hosaka N., Semba S., Shoji A., Sato M., Hamasaki M., Yuki S., Sano S., Segawa Y., et al. Fully Automated Molecular Diagnostic System “Simprova” for Simultaneous Testing of Multiple Items. Sci. Rep. 2020;10:5409. doi: 10.1038/s41598-020-62109-5. - DOI - PMC - PubMed
    1. Jonguitud-Borrego N., Malci K., Anand M., Baluku E., Webb C., Liang L., Barba-Ostria C., Guaman L.P., Hui L., Rios-Solis L. High-throughput and automated screening for COVID-19. Front. Med. Technol. 2022;4:969203. doi: 10.3389/fmedt.2022.969203. - DOI - PMC - PubMed
    1. Chang C.K., Jian M.J., Chung H.Y., Lin J.C., Hsieh S.S., Tang S.H., Perng C.L., Chen C.W., Hung K.S., Chang F.Y., et al. Clinical Comparative Evaluation of the LabTurbo(TM) AIO((R)) Reverse Transcription-Polymerase Chain Reaction and World Health Organization-Recommended Assays for the Detection of Emerging SARS-CoV-2 Variants of Concern. Infect. Drug Resist. 2022;15:595–603. doi: 10.2147/IDR.S349669. - DOI - PMC - PubMed
    1. Chung H.Y., Jian M.J., Chang C.K., Lin J.C., Yeh K.M., Chen C.W., Yang Y.S., Hsieh S.S., Chen E.S., Yang M.H., et al. Multicenter study evaluating novel multi-specimen pooling assay for the detection of SARS-CoV-2: High sensitivity and high throughput testing. Pt 1J. Microbiol. Immunol. Infect. 2022;55:1069–1075. doi: 10.1016/j.jmii.2021.08.003. - DOI - PMC - PubMed
    1. Chung H.Y., Jian M.J., Chang C.K., Lin J.C., Yeh K.M., Yang Y.S., Chen C.W., Hsieh S.S., Tang S.H., Perng C.L., et al. Multicenter study evaluating one multiplex RT-PCR assay to detect SARS-CoV-2, influenza A/B, and respiratory syncytia virus using the LabTurbo AIO open platform: Epidemiological features, automated sample-to-result, and high-throughput testing. Aging. 2021;13:24931–24942. doi: 10.18632/aging.203761. - DOI - PMC - PubMed