. 2023 Jun;18(6):1093-1099.
doi: 10.1007/s11548-023-02869-3. Epub 2023 Mar 30.

Robotically controlled three-dimensional micro-ultrasound for prostate biopsy guidance

Affiliations

Robotically controlled three-dimensional micro-ultrasound for prostate biopsy guidance

Reid Vassallo et al. Int J Comput Assist Radiol Surg. 2023 Jun.

Abstract

Purpose: Prostate imaging to guide biopsy remains unsatisfactory, with current solutions suffering from high complexity and poor accuracy and reliability. One novel entrant into this field is micro-ultrasound (microUS), which uses a high-frequency imaging probe to achieve very high spatial resolution, and achieves prostate cancer detection rates equivalent to multiparametric magnetic resonance imaging (mpMRI). However, the ExactVu transrectal microUS probe has a unique geometry that makes it challenging to acquire controlled, repeatable three-dimensional (3D) transrectal ultrasound (TRUS) volumes. We describe the design, fabrication, and validation of a 3D acquisition system that allows for the accurate use of the ExactVu microUS device for volumetric prostate imaging.

Methods: The design uses a motorized, computer-controlled brachytherapy stepper to rotate the ExactVu transducer about its axis. We perform geometric validation using a phantom with known dimensions and compare performance with magnetic resonance imaging (MRI) using a commercial quality assurance anthropomorphic prostate phantom.

Results: Our geometric validation shows accuracy of 1 mm or less in all three directions, and images of an anthropomorphic phantom qualitatively match those acquired using MRI and show good agreement quantitatively.

Conclusion: We describe the first system to acquire robotically controlled 3D microUS images using the ExactVu microUS system. The reconstructed 3D microUS images are accurate, which will allow for future applications of the ExactVu microUS system in prostate specimen and in vivo imaging.

Keywords: Micro-ultrasound; Prostate cancer; Robotics; Three dimensional.

PubMed Disclaimer

References

    1. Sung H, Ferlay J, Siegel RL, Laversanne M, Soerjomataram I, Jemal A, Bray F (2021) Global cancer statistics 2020: Globocan estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 71(3):209–249 - DOI - PubMed
    1. Izawa JI, Klotz L, Siemens DR, Kassouf W, So A, Jordan J, Chetner M, Iansavichene AE (2011) Prostate cancer screening: Canadian guidelines 2011. Can Urol Assoc J 5(4):235 - DOI - PubMed - PMC
    1. Onur R, Littrup PJ, Pontes JE, Bianco FJ Jr (2004) Contemporary impact of transrectal ultrasound lesions for prostate cancer detection. J Urol 172(2):512–514 - DOI - PubMed
    1. Serefoglu EC, Altinova S, Ugras NS, Akincioglu E, Asil E, Balbay D (2013) How reliable is 12-core prostate biopsy procedure in the detection of prostate cancer? Can Urol Assoc J 7(5–6):e293-8 - DOI - PubMed - PMC
    1. Wegelin O, Exterkate L, van der Leest M, Kummer JA, Vreuls W, de Bruin PC, Bosch JR, Barentsz JO, Somford DM, van Melick HH (2019) The future trial: a multicenter randomised controlled trial on target biopsy techniques based on magnetic resonance imaging in the diagnosis of prostate cancer in patients with prior negative biopsies. Eur Urol 75(4):582–590 - DOI - PubMed

LinkOut - more resources