The Fate of the Bent Rod in Children With Osteogenesis Imperfecta
- PMID: 37026790
- DOI: 10.1097/BPO.0000000000002409
The Fate of the Bent Rod in Children With Osteogenesis Imperfecta
Abstract
Background: Osteogenesis imperfecta (OI) is a genetic disorder characterized by brittle bones and long bone deformity. Realignment and intramedullary rodding with telescopic rods are indicated for progressive deformity and can help prevent fractures. Rod bending is a reported complication of telescopic rods and a common indication for revision; however, the fate of bent lower extremity telescopic rods in the setting of OI has not been reported.
Methods: Patients with OI at a single institution who underwent lower extremity telescopic rod placement with at least 1-year follow-up were identified. Bent rods were identified, and for these bone segments, we collected the location and angle of bend, subsequent telescoping, refracture, increasing angulation of bend, and date of revision.
Results: One hundred sixty-eight telescopic rods in 43 patients were identified. Forty-six rods (27.4%) bent during follow-up, with an average angulation of 7.3 (range: 1 to 24) degrees. In patients with severe OI, 15.7% of rods bent compared with 35.7% in nonsevere OI ( P =0.003). The proportion of bent rods was different between independent and nonindependent ambulators (34.1% and 20.5%; P =0.035). Twenty-seven bent rods (58.7%) were revised, with 12 rods (26.0%) revised early (within 90 d). The angulation of rods that were revised early was significantly higher than rods not (14.6 and 4.3 degrees, P <0.001). Of the 34 bent rods not revised early, the average time to revision or final follow-up was 29.1 months. Twenty-five rods (73.5%) continued to telescope, 14 (41.2%) increased in angulation (average 3.2 degrees), and 10 bones (29.4%) refractured. None of the refractures required immediate rod revision. Two bones had multiple refractures.
Conclusions: Bending is a common complication of telescopic rods in the lower extremities of patients with OI. It is more common in independent ambulators and patients with nonsevere OI, possibly because of the increased demand placed on the rods. Rods with a small bend and maintained fixation can telescope and need not be an indication for immediate revision.
Level of evidence: Level III-Retrospective review.
Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
The authors declare no conflicts of interest.
References
-
- Plotkin H. Two questions about osteogenesis imperfecta. J Pediatr Orthop. 2006;26:148–149.
-
- Franzone JM, Shah SA, Wallace MJ, et al. Osteogenesis imperfecta: a pediatric orthopedic perspective. Orthop Clin North Am. 2019;50:193–209.
-
- Ruck J, Dahan-Oliel N, Montpetit K, et al. Fassier-Duval femoral rodding in children with osteogenesis imperfecta receiving bisphosphonates: functional outcomes at one year. J Child Orthop. 2011;5:217–224.
-
- Nicolaou N, Bowe JD, Wilkinson JM, et al. Use of the Sheffield telescopic intramedullary rod system for the management of osteogenesis imperfecta: clinical outcomes at an average follow-up of nineteen years. J Bone Joint Surg Am. 2011;93:1994–2000.
-
- Montpetit K, Palomo T, Glorieux FH, et al. Multidisciplinary treatment of severe osteogenesis imperfecta: functional outcomes at skeletal maturity. Arch Phys Med Rehabil. 2015;96:1834–1839.
MeSH terms