Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2023 Mar 31;12(7):2626.
doi: 10.3390/jcm12072626.

Mechanical Insufflation-Exsufflation: Considerations for Improving Clinical Practice

Affiliations
Review

Mechanical Insufflation-Exsufflation: Considerations for Improving Clinical Practice

Michelle Chatwin et al. J Clin Med. .

Abstract

The provision of mechanical insufflation-exsufflation (MI-E) devices to enhance cough efficacy is increasing. Typically, MI-E devices are used to augment cough in patients with neuromuscular disorders but also in patients who are weak in an acute care setting. Despite a growing evidence base for the use of these devices, there are barriers to the provision of MI-E, including clinician lack of knowledge and confidence. Enhancing clinician education and confidence is key. Individualized or protocolized approaches can be used to initiate MI-E. Evaluation of MI-E efficacy is critical. One method to evaluate effectiveness of MI-E is the MI-E-assisted cough peak flow (CPF). However, this should always be considered alongside other factors discussed in this review. The purpose of this review is to increase the theoretical understanding of the provision and evaluation of MI-E and provide insight into how this knowledge can be applied into clinical practice. Approaches to initiation and titration can be selected based on the clinical situation, patient diagnosis (including and beyond neuromuscular disorders), and clinician's confidence.

Keywords: MI-E; airway clearance techniques; bulbar insufficiency; cough; cough assist; cough peak flow; neuromuscular disease.

PubMed Disclaimer

Conflict of interest statement

M.C. declares that she works part time for Breas Medical and is responsible for their non-commercial education and research portfolio. Breas Medical had no role in the preparation of this manuscript. Clinical the author holds honorary contracts to carry out clinical work at the NMCCC, the National Hospital for Neurology and Neurosurgery and the Respiratory Support Service at the Royal Brompton Hospital. R.H.W. declares no COI in relation to this manuscript.

Figures

Figure 1
Figure 1
Shows the components that occur within a normal cough. Blue represents the gastric pressure, red is the esophageal pressure, and black is the transdiaphragmatic pressure. Superimposed on this in grey are the flow changes that are seen during a cough. Arrow 1 indicates the start of the deep inspiration. Arrow 2 is the forced expiration against a closed glottis. Arrow 3 is the sudden opening of the glottis and Arrow 4, compression of the intra-thoracic airways. Cough peak flow (CPF) is measured as the peak expiratory flow that occurs during a cough. For anyone over the age of 12 years old, a CPF that exceeds 360 L/min is deemed to be normal. There are reference values for children [9].
Figure 2
Figure 2
Shows one approach to assessing the efficacy of a patient’s cough.
Figure 3
Figure 3
Shows a suggested algorithm for a standard approach to MI-E initiation in adult patients.
Figure 4
Figure 4
Shows a suggested algorithm for personalized approach to titration of mechanical insufflation-exsufflation (MI-E) devices in adult patients. This algorithm is modified from Chatwin and Simonds [57], taking into account more recent evidence around MI-E times and also MI-E-assisted cough peak flow (CPF). Inspiratory time (Ti), expiratory time (Te).
Figure 5
Figure 5
Shows pressure (red) and flow (blue) traces during a cycle of mechanical insufflation exsufflation with a patient coughing. The first negative flow point is artifact as it is the compressible volume. The second peak is the true cough peak flow (CPF). With thanks to Manuel Lujan.
Figure 6
Figure 6
Shows the flow, pressure, and inspiratory tidal volumes (Vti) download for a patient using an E70 (Philips Respironics, Murrysville, PA, USA). The blue line shows the device assisted cough peak flow (CPF) of around 350 L/min. Clinicians would normally be very pleased with this reading as it is at a level that would indicate an effective assisted CPF. However, after the initial CPF, there is flattening of the exsufflation curve, indicating an obstructed airway. With thanks to Tiago Pinto.

References

    1. Gross D., Zidulka A., O’Brien C., Wight D., Fraser R., Rosenthal L., King M. Peripheral mucociliary clearance with high-frequency chest wall compression. J. Appl. Physiol. 1985;58:1157–1163. doi: 10.1152/jappl.1985.58.4.1157. - DOI - PubMed
    1. Warwick W.J. Mechanisms of mucous transport. Eur. J. Respir. Dis. Suppl. 1983;127:162–167. - PubMed
    1. Camner P. Studies on the removal of inhaled particles from the lungs by voluntary coughing. Chest. 1981;80:824–827. - PubMed
    1. Lannefors L., Button B.M., McIlwaine M. Physiotherapy in infants and young children with cystic fibrosis: Current practice and future developments. J. R. Soc. Med. 2004;97((Suppl. S44)):8–25. - PMC - PubMed
    1. Chatwin M., Toussaint M., Gonçalves M.R., Sheers N., Mellies U., Gonzales-Bermejo J., Sancho J., Fauroux B., Andersen T., Hov B., et al. Airway clearance techniques in neuromuscular disorders: A state of the art review. Respir. Med. 2018;136:98–110. doi: 10.1016/j.rmed.2018.01.012. - DOI - PubMed

LinkOut - more resources