. 2023 Jul 1:259:124536.
doi: 10.1016/j.talanta.2023.124536. Epub 2023 Apr 11.

Lab-made 3D-printed electrochemical sensors for tetracycline determination

Affiliations

Lab-made 3D-printed electrochemical sensors for tetracycline determination

Carlos E C Lopes et al. Talanta. .

Abstract

Antibiotics such as tetracycline (TC) are widely prescribed to treat humans or dairy animals. Therefore, it is important to establish affordable devices in laboratories with minimal infrastructure. 3D printing has proven to be a powerful and cost-effective tool that revolutionizes many applications in electrochemical sensing. In this work, we employ a conductive filament based on graphite (Gr) and polylactic acid (PLA) (40:60; w/w; synthesized in our lab) to manufacture 3D-printed electrodes. This electrode was used "as printed" and coupled to batch injection analysis with amperometric detection (BIA-AD) for TC sensing. Preliminary studies by cyclic voltammetry and differential pulse voltammetry revealed a mass transport governed by adsorption of the species and consequent fouling of the redox products on the 3D printed surface. Thus, a simple strategy (solution stirring and application of successive potentials, +0.95 V followed by +1.2 V) was associated with the BIA-AD system to solve this effect. The proposed electrode showed analytical performance comparable to costly conventional electrodes with linear response ranging from 0.5 to 50 μmol L-1 and a detection limit of 0.19 μmol L-1. Additionally, the developed method was applied to pharmaceutical, tap water, and milk samples, which required minimal sample preparation (simple dilution). Recovery values of 92-117% were obtained for tap water and milk samples, while the content found of TC in the capsule was close to the value reported by the manufacturer. These results indicate the feasibility of the method for routine analysis involving environmental, pharmaceutical, and food samples.

Keywords: 3D-printing; Amperometry; Antibiotic; Batch injection analysis; Conductive filament.

PubMed Disclaimer

Conflict of interest statement

Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

LinkOut - more resources