Determination of the Re-188 Calibration Number for the Capintec CRC- 25PET Dose Calibrator
- PMID: 37138431
- DOI: 10.2174/1874471016666230502140224
Determination of the Re-188 Calibration Number for the Capintec CRC- 25PET Dose Calibrator
Abstract
Background: During the development of novel Re-188 radiopharmaceuticals, it was discovered that no calibration settings were published to calibrate Re-188 on the Capintec CRC-25PET dose calibrator.
Methods: Sodium [188Re]perrhenate was eluted from an OncoBeta 188W/188Re generator to measure activity on a Capintec CRC-25R dose calibrator using established dose calibrator settings provided by the manufacturer. The eluent was then used to tune the calibra on settings on a Capintec CRC-25PET dose calibrator, accounting for geometry. Radionuclidic purity of the [188Re]perrhenate source was verified via gamma spectroscopy.
Results: The calibrator number for Re-188 was determined to be 469 x 10 for the Capintec CRC-25PET dose calibrator, which differed from the manufacturer provided calibra on number of 496 x 10 for the Capintec CRC-25R dose calibra on model. W-188 breakthrough was characterised as < 0.01%.
Conclusion: This previously unreported calibration number can be used to determine the activity of Re- 188 labelled radiopharmaceuticals using the Capintec CRC-25PET dose calibrator model.
Keywords: Rhenium-188; [188Re]perrhenate; calibration numbers; dose calibrators; radiopharmaceuticals; theranostics.