Optimization of the atrioventricular delay in conduction system pacing
- PMID: 37161936
- DOI: 10.1111/jce.15927
Optimization of the atrioventricular delay in conduction system pacing
Abstract
Introduction: In patients receiving conduction system pacing (CSP), it is not well established how to program the sensed atrioventricular delay (sAVD), with respect to the type of capture obtained (selective, nonselective His-bundle [HB] capture or left bundle branch [LBB] capture). The aim of this study was to acutely assess the effectiveness of an electrophysiology (EP)-guided method for sAVD optimization by comparing it with the echocardiogram-guided optimization.
Methods and results: Consecutive patients undergoing HB or LBB pacing were enrolled. The EP-guided sAVD was defined as the sAVD leading to a PR interval of 150 ms on surface electrocardiogram (ECG). In HB pacing patients, EP-guided sAVD was obtained subtracting the time from the onset of the P wave on ECG to the local atrial electrogram (EGM) recorded by the atrial lead (right atrial sensing latency, RASL) and the His-ventricular interval from 150 ms; in LBB pacing patients, subtracting RASL from 150 ms. Transmitral flow assessment by pulsed wave Doppler was used to find the echo-optimized sAVD by a modified iterative method. The discordance between the EP-guided and the echo-optimized sAVD was recorded.
Results: Seventy-one patients were enrolled: 12 with selective, 32 nonselective HB capture, and 27 LBB capture. Overall, the rate of concordance between the EP-guided and the echo-optimized sAVD was 71.8%, with no significant differences between the three groups.
Conclusion: In CSP patients, an optimal sAVD can be programmed, in more than 70% of cases, considering only simple EGM intervals to obtain a physiological PR interval on surface ECG.
Keywords: AV delay optimization; His bundle pacing; conduction system pacing; left bundle branch area pacing.
© 2023 Wiley Periodicals LLC.
References
REFERENCES
-
- Deshmukh P, Casavant DA, Romanyshyn M, Anderson K. Permanent, direct His-bundle pacing: a novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation. 2000;101:869-877.
-
- Vijayaraman P, Naperkowski A, Subzposh FA, et al. Permanent His-bundle pacing: long-term lead performance and clinical outcomes. Heart Rhythm. 2018;15:696-702.
-
- Huang W, Su L, Wu S, et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can J Cardiol. 2017;33:1736.e1-1736.e3.
-
- Li Y, Chen K, Dai Y, et al. Left bundle branch pacing for symptomatic bradycardia: implant success rate, safety, and pacing characteristics. Heart Rhythm. 2019;16:1758-1765.
-
- Ravi V, Beer D, Pietrasik GM, et al. Development of new-onset or progressive atrial fibrillation in patients with permanent HIS bundle pacing versus right ventricular pacing: results from the RUSH HBP registry. J Am Heart Assoc. 2020;9:e018478.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials