Accuracy of implant site preparation in robotic navigated dental implant surgery
- PMID: 37199055
- DOI: 10.1111/cid.13224
Accuracy of implant site preparation in robotic navigated dental implant surgery
Abstract
Background: Modern technological advancements have led to increase in the development of surgical robots in dentistry, resulting in excellent clinical treatment outcomes.
Purpose: This study aimed to determine the accuracy of automatic robotic implant site preparation for different implant sizes by correlating planned and posttreatment positions, and to compare the performance of robotic and human freehand drilling.
Method: Seventy-six drilling sites on partially edentulous models were used, with three different implant sizes (Ø = 3.5 × 10 mm, 4.0 × 10 mm, 5.0 × 10 mm). The robotic procedure was performed using software for calibration and step-by-step drilling processes. After robotic drilling, deviations in the implant position from the planned position were determined. The angulation, depth, and coronal and apical diameters on the sagittal plane of sockets created by human and robotic drilling were measured.
Results: The deviation of the robotic system was 3.78° ± 1.97° (angulation), 0.58 ± 0.36 mm (entry point), and 0.99 ± 0.56 mm (apical point). Comparison of implant groups showed the largest deviation from the planned position for 5 mm implants. On the sagittal plane, there were no significant differences between robotic and human surgery except for the 5-mm implant angulation, indicating similar quality between human and robotic drilling. Based on standard implant measurements, robotic drilling exhibited comparable performance to freehand human drilling.
Conclusions: A robotic surgical system can provide the greatest accuracy and reliability regarding the preoperative plan for small implant diameters. In addition, the accuracy of robotic drilling for anterior implant surgery can also be comparable to that of human drilling.
Keywords: digital dentistry; implant drilled socket; implantology; radiographs; surgical; techniques.
© 2023 Wiley Periodicals LLC.
References
REFERENCES
-
- Carbone M, Goss E, Borione M, et al. Implant supported prostheses with bone system implant system: a retrospective study with follow-up period up to 13-years about 1021 fixtures. Minerva Stomatol. 2007;56(10):481-495.
-
- Karoussis IK, Brägger U, Salvi GE, Bürgin W, Lang NP. Effect of implant design on survival and success rates of titanium oral implants: a 10-year prospective cohort study of the ITI® dental implant system. Clin Oral Implants Res. 2004;15(1):8-17.
-
- Yan B, Zhang W, Cai L, et al. Optics-guided robotic system for dental implant surgery. Chin J Mech Eng. 2022;35(1):1-13.
-
- Brief J, Edinger D, Hassfeld S, Eggers G. Accuracy of image-guided implantology. Clin Oral Implants Res. 2005;16(4):495-501.
-
- Shan T, Tay F, Gu L. Application of artificial intelligence in dentistry. J Dent Res. 2021;100(3):232-244.
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources