Randomized Controlled Trial
. 2023 May 19;25(5):euad134.
doi: 10.1093/europace/euad134.

Effects of hyperbaric exposure on mechanical and electronic parameters of implantable cardioverter-defibrillators

Affiliations
Randomized Controlled Trial

Effects of hyperbaric exposure on mechanical and electronic parameters of implantable cardioverter-defibrillators

Romain Guenneugues et al. Europace. .

Abstract

Aims: Hyperbaric oxygen therapy (HBOT) is the standard adjuvant treatment for life-threatening or disabling pathologies. Currently, mechanical and electronic variations of implantable cardioverter-defibrillators (ICD) in hyperbaric conditions have not been evaluated. As a result, many patients eligible for HBOT but ICD recipients cannot undergo this therapy, even in emergency situations.

Methods and results: Twenty-two explanted ICD of various brands and models were randomized in two groups: single hyperbaric exposure at an absolute pressure of 4000 hPa and 30 iterative hyperbaric exposures at an absolute pressure of 4000 hPa. Mechanical and electronic parameters of these ICD were blindly assessed before, during, and after hyperbaric exposures. Regardless of the hyperbaric exposure, we could not find any mechanical distortion, inappropriate occurrence of anti-tachycardia therapies, dysfunction of tachyarrhythmia therapeutic programming, or dysfunction of programmed pacing parameters.

Conclusion: Dry hyperbaric exposure seems harmless on ICD tested ex vivo. This result may lead to a reconsideration of the absolute contraindication of emergency HBOT to ICD recipients. A real-life study in these patients with an indication to HBOT should be performed to assess their tolerance to the treatment.

Keywords: Hyperbaric exposure; Hyperbaric oxygen therapy; ICD.

PubMed Disclaimer

Conflict of interest statement

Conflict of interest: None declared.

Figures

Graphical abstract
Graphical abstract

References

    1. Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IMet al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Europace 2022;24:71–164. - PubMed
    1. Stühlinger M, Burri H, Vernooy K, Garcia R, Lenarczyk R, Sultan Aet al. EHRA Consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices. Europace 2022;24:1512–37. - PMC - PubMed
    1. Boersma LV, El-Chami M, Steinwender C, Lambiase P, Murgatroyd F, Mela Tet al. Practical considerations, indications, and future perspectives for leadless and extravascular cardiac implantable electronic devices: a position paper by EHRA/HRS/LAHRS/APHRS. Europace 2022;24:1691–708. - PubMed
    1. Mathieu D, Marroni A, Kot J. Tenth European Consensus Conference on Hyperbaric Medicine: recommendations for accepted and non-accepted clinical indications and practice of hyperbaric oxygen treatment. Diving Hyperb Med 2017;47:24–32. - PMC - PubMed
    1. Kratz JM, Blackburn JG, Leman RB, Crawford FA. Cardiac pacing under hyperbaric conditions. Ann Thorac Surg 1983;36:66–8. - PubMed

Publication types