Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 May;25(5):1098612X231171434.
doi: 10.1177/1098612X231171434.

Evaluation of a smartphone-based colorimetric method for urinalysis dipstick readings in cats

Affiliations

Evaluation of a smartphone-based colorimetric method for urinalysis dipstick readings in cats

Vincent Leynaud et al. J Feline Med Surg. 2023 May.

Abstract

Objectives: The aim of the study was to compare the diagnostic performances of a smartphone-based colorimetric method (SBCM) for urinalysis with a semi-automated point-of-care (POC) analyser using standardised solutions and cat urine.

Methods: Artificial solutions (negative and positive quality controls, and purposely designed artificial urine) and natural urine from 216 cats were used. Two urine reagent strips were simultaneously dipped in each sample. One dipstick was read by the SBCM and the other by the POC analyser at the same time. Results for pH, proteins, bilirubin, 'blood', glucose and ketones were considered. Overall agreement and sensitivity, specificity and accuracy of the SBCM were determined based on selected cut-offs.

Results: For the artificial solutions, 80 comparisons were obtained for each analyte and each expected concentration. The overall agreement (exactly the same result) between the two methods was 78.4%. SBCM sensitivity, specificity and accuracy were 99.0%, 100% and 99.3%, respectively. The correlation between the two methods was almost perfect (Cohen's kappa coefficient = 0.9851). For natural urine samples, the overall agreement (including pH) was 68.6%. Using optimal cut-offs for the SBCM determined from the results of analysis of artificial solutions, the sensitivity, specificity and accuracy of the SBCM were 100%, 76.02% and 80.5%, respectively. In this situation, the correlation between the two methods was moderate (Cohen's kappa coefficient = 0.5401). This was mostly due to a high rate of false-positive results for bilirubin (61.1%).

Conclusions and relevance: With proper cut-off use (ie, considering positive or negative results) the SBCM evaluated here has a perfect sensitivity and appropriate diagnostic performances for proteins, 'blood', glucose and ketones. Based on these experimental data, this method appears suitable for dipstick urinalysis but positive results for bilirubin and proteins have to be confirmed.

Keywords: Urinalysis; app; colorimetry; dipstick; smartphone.

PubMed Disclaimer

Conflict of interest statement

The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Figures

Figure 1
Figure 1
Percentages of identical and different clinical interpretations for each analyte from the results obtained with the smartphone-based colorimetric method and the point-of-care analyser from 216 cat urine samples
Figure 2
Figure 2
Bubble chart comparing pH results obtained by the smartphone-based colorimetric method (SBCM) and the point-of-care (POC) analyser in 216 cat urine samples. Circle sizes are proportional to the number of results (reported within or near corresponding circles)

References

    1. Moses JC, Adibi S, Shariful Islam SM, et al.. Application of smartphone technologies in disease monitoring: a systematic review. Healthcare 2021; 9: 889. DOI: 10.3390/healthcare9070889. - DOI - PMC - PubMed
    1. Liu J, Geng Z, Fan Z, et al.. Point-of-care testing based on smartphone: the current state-of-the-art (2017-2018). Biosens Bioelectron 2019; 132: 17–37. - PubMed
    1. Zhang M, Cui X, Li N. Smartphone-based mobile biosensors for the point-of-care testing of human metabolites. Mater Today Bio 2022; 14. DOI: 10.1016/j.mtbio.2022.100254. - DOI - PMC - PubMed
    1. Krimer PM, Tanner MC, Camus MS. Evaluation of a home urinalysis kit in dogs. J Am Anim Hosp Assoc 2019; 55: 144–151. - PubMed
    1. Peele JD, Jr, Gadsden RH, Crews R. Semi-automated vs. visual reading of urinalysis dipsticks. Clin Chem 1977; 23: 2242–2246. - PubMed

Publication types

LinkOut - more resources