Rice-Body Synovitis, Foreign Body Reaction, and Rotator Cuff Failure After Subacromial Balloon Spacer Augmentation of a Rotator Cuff Repair: A Case Report
- PMID: 37235697
- DOI: 10.2106/JBJS.CC.23.00009
Rice-Body Synovitis, Foreign Body Reaction, and Rotator Cuff Failure After Subacromial Balloon Spacer Augmentation of a Rotator Cuff Repair: A Case Report
Abstract
Case: A 66-year-old woman presented with shoulder pain and weakness 4 months after augmentation of a rotator cuff repair with a Stryker InSpace subacromial balloon spacer. A magnetic resonance imaging (MRI) demonstrated a failed rotator cuff repair, large effusion with rice bodies, synovitis, axillary lymphadenopathy, loose anchors, and erosive changes to the greater tuberosity. Arthroscopy revealed balloon fragmentation surrounded by diffusely hyperemic synovium without repairable cuff tissue. Final cultures proved negative for infection. Histologic evaluation revealed ulcerated synovium with diffuse chronic and focal acute inflammation.
Conclusion: Despite promising early results, augmentation of a rotator cuff repair with a subacromial balloon spacer introduces a risk of inflammatory reaction that may mimic a deep infection and compromise rotator cuff healing.
Copyright © 2023 by The Journal of Bone and Joint Surgery, Incorporated.
Conflict of interest statement
Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSCC/C101).
References
-
- Verma N, Srikumaran U, Roden CM, Rogusky EJ, Lapner P, Neill H, Abboud JA. InSpace implant compared with partial repair for the treatment of full-thickness massive rotator cuff tears: a multicenter, single-blinded, randomized controlled trial. J Bone Joint Surg Am. 2022;104(14):1250-62.
-
- Metcalfe A, Parsons H, Parsons N, Brown J, Fox J, Gemperlé Mannion E, Haque A, Hutchinson C, Kearney R, Khan I, Lawrence T, Mason J, Stallard N, Underwood M, Drew S, Metcalfe A, Parsons H, Parsons N, Arif A, Arnold S, Bhabra G, Brown J, Bora S, Bush H, Drew S, Fox J, Gemperlé Mannion E, Haque A, Hutchinson C, Jones C, Kearney R, Khan I, Lawrence T, Mason J, McGowan K, Modi C, Rahman B, Rahman U, Ramirez M, Spocinska M, Stallard N, Teuke J, Thayalan V, Ul-Rahman S, Viswanath A, Underwood M, Haque A, Khan I, Mason J, Parsons H, Parsons N, Stallard N, Packham I, Barnett E, Witham R, Crowther M, Murphy R, Coates K, Morley J, Barnfield S, Gill S, Jones A, Halliday R, Dunn S, Fagg J, Dacombe P, Nanda R, Wilson D, Boulton L, Liow R, Jeavons R, Meddes A, Kang N, Dehghani L, Nacorda A, Punnoose A, Ferran N, Adewetan G, Adedoyin T, Pall A, Sala M, Zaman T, Hartley R, a-Sayyad C, Vamplew L, Howe E, Bokor N, Corbett S, Moverley R, Cox E, Donaldson O, Jones M, Wood D, Perry J, Lewis A, Howard L, Beesley K, Harries L, Elmorsy A, Wilcocks K, Shean K, Diment S, Pidgeon H, King V, Sjolin S, Williams A, Kellett J, Young L, Dunne M, Lockwood T, Curtis M, Siddiqui N, Mckenley I, Morrison S, Morrison C, O'Brien T, Bradley I, Lambatan K, Kelly C, Perkins C, Jones T, Rowlands T, Collins D, Nicholas C, Birch C, Lloyd -Evans J, Akhbari P, Edakalathu JJ, Hand C, Cole A, Prince D, Thorpe K, Rooke L, Baggot M, Morris M, Ivanova D, Baker D, Matthews T, Falatoori J, Jarvis H, Jones D, Williams M, Evans R, Pullen H, Hodkinson G, Vannet N, Davey A, Poyser E, Hall A, Mehta H, Prakash Tokola D, Connor C, Jordan C, Ennis O, Omar Z, Lewis T, Owen AL, Morgan A, Ponnada R, Al-Azzani W, Williams C, Knox L, Singh H, Lee T, Robinson K, Kumar D, Armstrong A, Majed A, Falworth M, Butt D, Higgs D, Rudge W, Hughes B, Hanison E, Brooking D, Patel A, Symonds A, Gibson J, Santiago R, Barlow D, Lennon J, Smith C, Hall J, Griffin E, Lear R, Thomas W, Rose D, Edkins J, Samuel H, Jahnich H, Geoghegan J, Gooding B, Hudson S, Nightingale J, Papanna M, Briggs T, Pugh R, Neal A, Warrem L, Maxwell V, Chadwick R, Jaki T, Davies L, Gwilym S, Taylor R, Abel G, Graham J, Littlewood C, Wallace A, Howard A. Subacromial balloon spacer for irreparable rotator cuff tears of the shoulder (START:REACTS): a group-sequential, double-blind, multicentre randomised controlled trial. Lancet. 2022;399(10339):1954-63.
-
- Garofalo R, De Crescenzo A, Fontanarosa A, Conti M, Castagna A, Calbi R. Rotator cuff repair protected with subacromial balloon spacer shows a low rate of non-healing. Knee Surg Sports Traumatol Arthrosc. 2022;30(6):2123-9.
-
- Garrigues GE, Zmistowski B, Cooper AM, Green A, Hsu J, Ricchetti E, Namdari S, Frankle M, Gerber C, Tashjian R, Matsen F. Proceedings from the 2018 international consensus meeting on orthopedic infections: the definition of periprosthetic shoulder infection. J Shoulder Elbow Surg. 2019;28(6):S8-S12.
-
- Ramot Y, Nyska A, Markovitz E, Dekel A, Klaiman G, Zada MH, Domb AJ, Maronpot RR. Long-term local and systemic safety of poly(L-lactide-co-epsilon-caprolactone) after subcutaneous and intra-articular implantation in rats. Toxicol Pathol. 2015;43(8):1127-40.